Installation
Installation setups
Thermo Fisher Scientific
Thermo Scientific KingFisher Flex User Manual 33
Warning
This product contains very strong permanent magnets. People
wearing a pacemaker or metallic prostheses should not use this product. A
pacemaker or prostheses may be affected or damaged if it comes in close
contact with a strong magnetic field.
s
To insert KingFisher Flex heads, choose the Change_magnet protocol
under the Maintenance menu (Maintenance > Maintenance protocol >
Change_magnet) by using the up and down cursor keys. Then press
START
. Insert the KingFisher Flex head so that the three pins (shown with
black arrows) slot into place (Figure 3
–23 and Figure 3–24). When you
insert the KingFisher Flex head, be careful not to damage the magnet rods
against the tip comb holder frame. Confirm the end of the protocol by
pressing
START
.
Figure 3
–23.
Inserting the KingFisher Flex head