Installation
Operational check
Thermo Fisher Scientific
Thermo Scientific KingFisher Duo User Manual
29
4.
Take away the magnet head by turning the two finger screws on top
of the magnet head clockwise when instructed (Figure 3–15). When
you remove the KingFisher Duo head, be careful not to damage the
magnet rods against the tip comb holder frame.
Figure 3–15.
Removing the two finger screws on top of the magnet head
5.
Insert the other magnet head by slotting it into the tip comb holder
and by refitting the two finger screws on top of the magnet head
(Figure 3–16 and Figure 3–17).
6.
The tip moves on top of the shield plate at the end of the changing
process.
7.
Screw the maintenance door on again by screwing the two
red
screws on counterclockwise.
▲
Figure 3–16.
Inserting the KingFisher Duo head
Содержание KingFisher Duo
Страница 1: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 ...
Страница 2: ......
Страница 3: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 Cat no N12420 ...
Страница 12: ...Contents 10 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 22: ...Functional Description Reagents 20 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 74: ...Certificate of Decontamination 72 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 76: ...System Log 74 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 78: ...Glossary 76 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 81: ...Thermo Fisher Scientific Thermo Scientific KingFisher Duo User Manual 79 Notes ...
Страница 82: ...80 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific Notes ...
Страница 83: ......
Страница 84: ...Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com N12420 ...