Routine Operation
How to start
Thermo Fisher Scientific
Thermo Scientific KingFisher Duo User Manual
53
Caution
Only use plates that are listed in Table 4–4. Other plates may
not be compatible with the KingFisher Duo heating blocks. They may
also cause unexpected problems, such as cross-contamination due to the
divergent well volume and bottom height of the plate.
▲
Table 4–4.
Processing volumes vs. plate types and magnet heads
Head / Plate
KingFisher Flex 24 deep
well plate
Microtiter deep well
96 plate
KingFisher Duo
elution strip
KingFisher Duo head for
24 DW plate
200–5000 μl
–
–
KingFisher Duo head for
96 DW plate
–
50–1000 μl
30–130 μl
The KingFisher Duo elution strip can only be used with a KingFisher
Duo instrument installed with a 12-pin magnet head (Figure 4–28).
Figure 4–28.
KingFisher Duo elution strips, separate and in a frame
The elution strip is typically used to perform the elution step that
concentrates the purified sample. It is easy to pipette small volumes into
the elution strip due to its low design. The eight elution strips are also
convenient for storage and can be collected onto the 96 elution frame
(Figure 4–28).
To start the instrument:
1.
Select a protocol by using the arrow keys and press
START
or use
BindIt Software to run the desired protocol via the PC.
2.
Ensure that the front lid is open (Figure 2–3).
3.
Load the plate(s) and elution strip(s) in the order that the protocol
requests (Figure 4–30). Place the A1 well of the plate ensuring that
it is in the upper right corner. The first A1 row is consequently
always in the inner circle. Once you have loaded the requested
plates into their respective plate positions, press
OK
. The instrument
Elution strip
How to start
Содержание KingFisher Duo
Страница 1: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 ...
Страница 2: ......
Страница 3: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 Cat no N12420 ...
Страница 12: ...Contents 10 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 22: ...Functional Description Reagents 20 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 74: ...Certificate of Decontamination 72 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 76: ...System Log 74 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 78: ...Glossary 76 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 81: ...Thermo Fisher Scientific Thermo Scientific KingFisher Duo User Manual 79 Notes ...
Страница 82: ...80 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific Notes ...
Страница 83: ......
Страница 84: ...Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com N12420 ...