Maintenance
Packing for service
60
Thermo Scientific KingFisher Duo User Manual
Thermo Fisher Scientific
9.
Remove the instrument from the bag.
10.
Clean the instrument using a mild detergent.
11.
Remove any stains with 70% ethanol.
12.
After performing this decontamination procedure, enclose a signed
and dated Certificate of Decontamination inside the transport
package and attached to the outside of the package (see Appendix A:
“
Certificate of Decontamination
”).
To pack for service, follow the guidelines presented below.
Caution
It is important that the instrument is thoroughly
decontaminated before it is removed from the laboratory or any
servicing is performed on it.
▲
When you ship the instrument for service, remember to:
●
Inform about the use of hazardous materials.
●
Decontaminate the instrument beforehand.
●
Install the transport lock.
●
Place the KingFisher Duo head into its transport package.
●
Pack the instrument according to the enclosed packing instructions.
●
Use the original packaging to ensure that no damage will occur to
the instrument during shipping. Any damage will incur additional
labor charges.
●
Enclose a dated and signed Certificate of Decontamination (see
Appendix A: “
Certificate of Decontamination
”) inside and attached
to the outside of the package, in which you return your instrument
(or other items).
●
Enclose the return goods authorization number (RGA) given by
your Thermo Fisher Scientific representative.
●
Indicate the fault after you have contacted your local Thermo Fisher
Scientific representative or Thermo Fisher Scientific’s technical
service department.
Refer to “General specifications” on page 63 for details on storage and
transportation temperatures.
Packing for
service
Содержание KingFisher Duo
Страница 1: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 ...
Страница 2: ......
Страница 3: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 Cat no N12420 ...
Страница 12: ...Contents 10 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 22: ...Functional Description Reagents 20 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 74: ...Certificate of Decontamination 72 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 76: ...System Log 74 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 78: ...Glossary 76 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Страница 81: ...Thermo Fisher Scientific Thermo Scientific KingFisher Duo User Manual 79 Notes ...
Страница 82: ...80 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific Notes ...
Страница 83: ......
Страница 84: ...Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 FI 01621 Vantaa Finland www thermoscientific com N12420 ...