Clean the cell counter
Introduction
Clean the Countess
™
Automated Cell Counter periodically to prevent buildup of dust and dirt that might
reduce its performance and cause contamination.
CAUTION!
To avoid electrical shock, always turn OFF the Countess
™
Automated Cell Counter and
unplug the power cord before cleaning or decontaminating the instrument.
CAUTION!
All biological samples and materials that come into contact with them have the
potential to transmit infectious diseases and are considered biohazardous. Follow all applicable
local, state/provincial, and/or national regulations. Wear appropriate protective eyewear, clothing, and
gloves.
IMPORTANT!
Using a cleaning or decontaminating method other than that specified by the
manufacturer may result in damage to the instrument.
Clean the touchscreen
• Wipe the touchscreen of the Countess
™
Automated Cell Counter using a soft, lint-free cloth
moistened with an LCD cleaning solution. Do not apply excessive force during cleaning. Wipe
the touchscreen dry immediately after cleaning.
• Ensure that the cleaning solution does not enter the power button, the power inlet, the slide port, or
the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical shock when the
instrument is plugged in.
• Do not use abrasive cleaning solutions or material to prevent the touch-screen from getting
scratched.
Clean the instrument case
• Wipe the instrument case of the Countess
™
Automated Cell Counter using a soft, lint-free cloth
moistened with distilled water. Wipe the instrument dry immediately after cleaning.
• Ensure that water or other cleaning solutions do not enter the power button, the power inlet, the
slide port, or the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical shock when the
instrument is plugged in.
Chapter 9
Maintenance
Clean the cell counter
9
40
Countess
™
3 Automated Cell Counter User Guide