Creating and Editing Labware Files
Labware File Syntax
8-4
ID Scribe User Manual
Thermo Scientific
The starting position is the location that the ID Scribe will write the first
line of the first tube. By modifying the x starting position and the y
starting position, you can modify where all of the labware will be printed
by shifting all of the printing locations.
For example, suppose the x starting position is set in a Labware File as -
1,800 pixels to the left of the home position of the ID Scribe and you
notice that the writing is too far the left of all the tubes as in Figure 8-3.
Figure 8–3.
Too far left
If you were then to change the x starting position to -1760 (1,760 pixels
to the left of home), then all of the tubes would be shifted to the right.
Figure 8–4.
Too far right
Then an x starting position between -1760 and -1800 might be the
appropriate position to print in the center of the tubes.
X Starting
Position