Before Use
Setting up
Thermo Scientific
Cryofuge 5500i
2-3
4. Screw the threaded rods with the threaded nuts as far up as possible. Otherwise they might get
stuck when rolling the centrifuge off the pallet.
5. Remove the locking stabilizers.
6. Check that the steering casters point to the back.
7. Use several people and other means in order to roll the centrifuge off the pallet.
8. Once the centrifuge is in its final position mount the enclosed locking stabilizers.
9. Screw the threaded nuts down.
10. Place the locking stabilizers centrally under the threaded rods and align the centrifuge. All the
locking stabilizers should carry uniformly (check by a tilt test). The casters are some 5-15 mm off
the ground.
Threaded rod
Lock nut
Safety bracket
Locking stabilizer
Threaded nut
Threaded nut
WARNING
Due to the centrifuges weight it must be controlled by several people when
rolling of the pallet. Do not stand in front of the centrifuge when rolling down for the
centrifuge might run you over.