5
Maintenance and Care
Service of Thermo Fisher Scientific
Thermo Scientific
Cryofuge 5500i
5-5
4.
Release the rotor.
5.
Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
6.
Remove the centrifuge tubes and adapters and dispose of them or disinfect them.
7.
Rinse the centrifuge first with ethanol and then with de-ionized water.
•
Adhere strictly to the given application times.
8.
Wipe off the disinfectant.
9.
Rinse thoroughly with water.
10.
Dispose of
the decontamination solution according to the applicable guidelines.
Service of Thermo Fisher Scientific
Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an
authorized service technician. The service technicians check the following:
•
the electrical equipment
;
•
the suitability of the set-up site
;
•
the door lock and the safety system
;
•
the rotor
;
•
the fixation of the rotor and the drive shaft
.
Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are
performed for free during the warranty period and afterwards for a charge.
This is only valid
,
if the centrifuge has only been maintained by a Thermo Fisher Scientific service
technician.