Problem Solving
User Calibration Trouble Shooting.
8-4
G100 Range CO
2
Analyser
Thermo Scientific
User Calibration Trouble Shooting.
Cross-Gas Effects
Carbon dioxide is measured by infrared absorption at a wavelength specific to carbon dioxide.
Therefore, the carbon dioxide reading will not be affected by any other gases. The oxygen sensor is a
galvanic cell type and suffers virtually no influence from CO
2
, CO, H
2
S, NO
2
, SO
2
or H
2
, unlike
many other types of oxygen cell.
Hardware Reset
If for any reason the instrument ‘locks up’ and will not switch off, it is possible to force a hardware
reset. Press and hold the ‘On/Off ’ key for 10 seconds; wait at least 15 seconds and the instrument
should restart automatically.
Error
Remedy
User Zero failed
A possible reason for this is because the instrument is trying
to zero to a level which is outside the predetermined range
set when the unit was calibrated at the factory. To rectify
this, first ensure the unit contains absolutely none of the gas
which is being zeroed by flushing thoroughly with nitrogen.
If it will not zero, then refer to the instructions given in the
section. If the instrument continues
to fail to zero then the unit must be returned to the manufac-
turer for investigation.
Calibration failed
Check the span target is set to the correct value, if not, cor-
rect and retry spanning the channel. Repeat the entire proce-
dure, including zeroing the channel and then calibrate the
span. Ensure the reading is stable before spanning the chan-
nel.
Note
Performing a hardware reset may cause loss or corruption of currently stored data
including the time/date.
Содержание G100 Series
Страница 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Страница 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 14: ...Instrument Features Instrument Connection Points 4 4 G100 Range CO2 Analyser Thermo Scientific ...
Страница 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Страница 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 50: ...Important Notice to All Customers WEEE COMPLIANT 14 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Страница 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......