Calibration
Last Field Calibration
7-4
G100 Range CO
2
Analyser
Thermo Scientific
4. The instrument will then indicate a successful zero has been completed. Press the soft-key
‘Accept’ to confirm the calibration and ‘Store’ the new user offset.
Alternatively, soft-key ‘Reject’ to exit without change.
Span CO
2
Channel:
It is recommended that the instrument is spanned to target the desired reading range (e.g. 5%);
ideally this should not be a low level close to zero.
1. If not already preset, enter the
span target
, i.e. certified concentration of your calibration gas.
Press ‘Key 1’ and enter the new value. Then attach the gas and open regulator valve to allow the
gas to flow.
2. Press the ‘Start’ key and wait for the reading to stabilise. This can take a couple of minutes. Press
the ‘Pump’ key to draw sample gas.
3. Once a stable reading is shown press the soft-key ‘Accept’. A successful span calibration message
will then be displayed. Press the soft-key ‘Accept’ again to confirm the calibration and ‘Store’ the
new user span. Alternatively, press soft-key ‘Reject’ to exit without change.
Reset Factory Settings
This option will reset the instrument to its factory programmed calibration characteristics and will
clear the user calibration points for both gas channels.
1. To reset to factory settings, press ‘Key 1 - Factory Reset’ from the ‘Calibration’ menu.
2. To prevent the user calibration data being accidentally erased the user must confirm the action
by pressing the soft-key ‘Accept’, or soft-key ‘Reject’ to exit without change.
Last Field Calibration
This data can be found in the ‘Information’ screen accessed via the ‘Utilities’ menu. This option
displays the date that the last field calibration was performed on the instrument.
Note
If the calibration failed then purge and try again or select a different air source.
Note
If using G110 – very high concentrations of CO
2
may take up to 30 minutes to
purge completely.
Note
If the calibration failed then try again using a longer purge time or different target
gas.
User Calibration - Reset
Содержание G100 Series
Страница 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Страница 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 14: ...Instrument Features Instrument Connection Points 4 4 G100 Range CO2 Analyser Thermo Scientific ...
Страница 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Страница 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 50: ...Important Notice to All Customers WEEE COMPLIANT 14 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Страница 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......