General Operational Instructions
Warning and Error Codes
5-10
G100 Range CO
2
Analyser
Thermo Scientific
Once the reading memory is full it is not possible to store any more readings. When full and the
‘Store’ key is pressed or data logging is activated the instrument will show a brief message stating that
the memory is full and that no further data will be recorded.
1. From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel.
2. Press ‘Key 4’ to view data.
3. To clear the readings press the soft-key ‘More’ followed by ‘Key 1 - Delete all’.
Diagnostics
The ‘Diagnostics’ option enables remote technical support to identify and resolve issues with the
instrument and readings. If required, the operator may be asked to confirm the diagnostics displayed.
1. From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel.
2. Press ‘Key 5’ to view diagnostics.
Warning and Error Codes
When switched on the instrument will perform a predetermined self-test sequence taking
approximately 10 seconds. During this time many of the instrument’s working parameters and
settings are checked. If any operational parameters are out of specification or if the pre-programmed
recommended calibration/service date has passed, errors or warnings may be displayed.
Battery/Charging
The battery used in the instrument is 2 Ah Lithium-Ion cell. The instrument must be charged using
the power supply supplied with your instrument.
When plugged into the charger the instrument will power on and display charging. When complete
the display will change to show that the instrument is charged. To switch the instrument ON whilst
charging or charged is displayed the operator will need to switch the instrument off and then on
again.
When the instrument is already powered on, the operation is slightly different as the battery icon
changes to a flashing plug symbol. This will stop flashing when the charge is complete.
Note
For further information contact your equipment supplier.
Note
For further information please refer to section
of
this manual.
Note
Although the instrument can be powered via the USB connector it cannot be
charged via USB.
Note
A full charge will take approximately
3 hours
. Typically, a fully charged battery will
last 8-10 hours.
Содержание G100 Series
Страница 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Страница 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 14: ...Instrument Features Instrument Connection Points 4 4 G100 Range CO2 Analyser Thermo Scientific ...
Страница 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Страница 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 50: ...Important Notice to All Customers WEEE COMPLIANT 14 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Страница 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......