Calibration
Calibration Method
Thermo Scientific
G100 Range CO
2
Analyser
7-3
Calibration Method
Before you begin ensure the unit is stabilised at its working temperature before performing any of the
calibration operations.
To achieve the processes set out in this section, press ‘Key 3 – Calibration’ from the ‘Main’ menu. The
first screen displayed provides the option to select the gas that requires calibration.
The exact calibration method can vary depending on the gases used.
Zero CO
2
Channel:
For maximum accuracy it is recommended that the CO
2
Channel is zeroed using bottled gas (certified
100% N2). However, if nitrogen gas is not available the optional soda lime filter kit can be fitted to the
gas inlet. This allows the user to perform a zero using normal air as the soda lime filter will absorb
virtually all CO
2
from the sample air. For both these options select 'Key 1-Zero with N2' from the user
calibration menu.
If neither of the recommended methods is available the user can select the option to perform an air
calibration. This option assumes that the user has access to fresh air at around 390ppm. Generally, this
can be found outside or in a well ventilated corridor (typically, an office or lab would have a higher
CO
2
concentration).
1. From the ‘Calibration’ Menu, press ‘Key 2 - CO
2
channel’.
2. Press either ‘Key 1 - Zero with N2’ (recommended) or ‘Key 2 - Zero with Air’ fromthe menu.
Then, either attach the 100% N2 or sample pipe to allow access to freshair.
3. Press the ‘Start’ key. The instrument will now wait (approximately 60 seconds) for the gas reading
to stabilise at the correct level. If zeroing with air press the ‘Pump’ key to draw in fresh air.
Calibration
User Calibration
Содержание G100 Series
Страница 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Страница 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 14: ...Instrument Features Instrument Connection Points 4 4 G100 Range CO2 Analyser Thermo Scientific ...
Страница 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Страница 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 50: ...Important Notice to All Customers WEEE COMPLIANT 14 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Страница 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......