Thermo Scientific
G100 Range CO
2
Analyser
5-1
5
General Operational Instructions
Switching the Instrument On
1. To switch on the instrument, press the ‘On/Off ’ key briefly. There will be a short beep and a slight
pause followed by the Geotech logo.
2. The power on self-test will then commence (approximately 15 seconds) including warm-up time.
3. Assuming there are no warnings to display the instrument will continue to the ‘Main Read Screen’.
Switching the Instrument Off
1. Purge with fresh air. Run the pump for approximately 30 seconds or until the readings have
returned to normal levels.
2. To switch off the instrument, press the ‘On/Off ’ key briefly.
Instrument Main Read Screen
After the analyser has been switched on and the warm-up self-tests completed the analyser will display
the ‘Main Read Screen’.
Soft-Keys:
Peak - Enables the operator to display the peak reading.
Store - Enables the operator to store the reading for viewing/download later.
Hold - Enables the operator to hold the current reading being taken.
Note
Before the instrument is switched off a clean air purge should be performed. This
ensures that the instrument is free from gas and ready for the next measurement. This final
purge is especially important for the oxygen sensor as it may degrade if stored when
contaminated with gas.
Note
If the ‘Auto Off ’ utilities setting is set to ‘Yes’, the analyser will switch off
automatically after 10 minutes if not in use.
Pump (On)
Indicator Symbol
Main Read Screen
Soft-Keys
Battery Power
Indicator
Содержание G100 Series
Страница 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Страница 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 14: ...Instrument Features Instrument Connection Points 4 4 G100 Range CO2 Analyser Thermo Scientific ...
Страница 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Страница 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 50: ...Important Notice to All Customers WEEE COMPLIANT 14 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Страница 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......