Thermo Scientific
G100 Range CO
2
Analyser
8-1
8
Problem Solving
This section outlines various warning and error messages which the operator may receive during
general operation of the instrument. For further assistance please contact your equipment supplier.
Warnings and Errors
When switched on the instrument will perform a predetermined self-test sequence taking
approximately 15 seconds. During this time many of the instrument’s working parameters and settings
are checked.
If any operational parameters are out of specification or if the pre-programmed recommended
calibration/service date has passed, errors or warnings may be displayed.
Use the ‘Scroll Up’ and ‘Scroll Down’ keys to move through the list if required.
There are two types of warning that may be displayed:
• General warnings that may not affect the instrument’s function and those where the self-test has
detected a function that is outside the usual operating criteria, e.g. battery charge low, memory
nearly full.
• Operational parameters that could affect the performance of the instrument, e.g. CO
2
out of
calibration.
The most likely reason for these errors is either an incorrect user calibration, or may indicate sensor
failure. If an incorrect user calibration has caused the warning it should be correctable by way of
returning the instrument to factory settings, zeroing or carrying out a user calibration as necessary for
the relevant function.
Under and Over Range Codes
If a reading is over range (i.e. above the maximum allowed reading) it will be displayed with more than
chevrons (>>.>). This can occur if a channel has been incorrectly calibrated or the sample gas has
exceeded its specified range (e.g. CO
2
> 20%).
If a reading is under range (i.e. below zero) it will be displayed with less than chevrons (<<. <). Refer to
section
of this manual to remedy under-range by performing a user zero.
A number displayed as asterisks (**.*) indicates an error, usually where the instrument has been unable
to complete a particular calculation. Typically, this will be the first indication of a fault condition.
Where no data is available dashes (--.-) are displayed. This usually occurs when a particular reading or
parameter has been skipped by the user, or where an optional accessory is not fitted correctly, i.e. a
temperature probe.
Содержание G100 Series
Страница 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Страница 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 14: ...Instrument Features Instrument Connection Points 4 4 G100 Range CO2 Analyser Thermo Scientific ...
Страница 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Страница 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 50: ...Important Notice to All Customers WEEE COMPLIANT 14 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Страница 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......