18
Thermo Scientifi c Novus SCP/MCP User Manual
Thermo Fisher Scientifi c
TABLE1: Maximum permissible errors according ISO 8655
Two point calibration
Normally the two-point calibration should be used.
1. Do the pipetting series with both max. and min. volumes.
2. Calculate the actual results.
3. Press MENU and select options with SCROLLKEY and accept with OK.
4. Select CALIBRATE and press OK.
5. Select two points and confi rm with OK. The target max. and target min.
volumes are shown on the display.
6. Press EDIT and change the actual max. volume with the SCROLLKEY and
accept with OK.
7. Change the actual min. volume with the SCROLLKEY and accept with OK.
8. A text SAVE? appears on the display.
9. Accept with YES
10. The adjustment has been changed.
Adjustment
Range Volume
Inaccuracy
Imprecision
µl
µl
%
s.d.
µl
cv%
1-10 µl
10
±0.120
±1.2
0.080
0.8
1
±0.120
±12
0.080
8.0
5-50 µl
50
±0.50
±1.0
0.20
0.4
5
±0.50
±10
0.20
4.0
10-100 µl
100
±0.80
±0.8
0.30
0.3
10
±0.80
±8.0
0.30
3.0
30-300 µl
300
±4.0
±1.3
1.5
0.5
30
±4.0
±13
1.5
5.0
100-1000 µl
1000
±8.0
±0.8
3.0
0.3
100
±8.0
±8.0
3.0
3.0
0,5-5 ml
5000
±40.0
±0.8
15.0
0.3
500
±40.0
±8.0
15.0
3.0
1-10 ml
10000
±60.0
±0.6
30.0
0.3
1000
±60.0
±6.0
30.0
3.0
Range Channel
Volume Inaccuracy
Imprecision
µl
µl
%
s.d.µl
cv%
1-10 µl
8, 12
10
±0.24
±2.4
0.16
1.6
1
±0.24
±24
0.16
16
5-50 µl
8, 12, 16
50
±1.0
±2.0
0.4
0.8
5
±1.0
±20
0.4
8.0
30-300 µl
8, 12
300
±8.0
±2.7
3.0
1.0
30
±8.0
±26.7
3.0
10.0
100-1200 µl
8
1200
±32
±2.7
12
1.0
100
±32
±32
12
12