12
Thermo Scientifi c Novus SCP/MCP User Manual
Thermo Fisher Scientifi c
With STEPPER function repeated dispensing of selected volume is possible.
Choose the STEPPER function as described above.
Select the pipetting volume simply by pressing the scroll key up or down. The
display shows always the maximum number of steps during volume selection.
Accept the volume with OK. Next choose the number of steps with scroll key
and accept with OK.
Optionally select the speed. Press SPEED and the speed in starts to blink. Select
the speed with scroll key and accept with OK. Select the speed out with scroll
key and accept with OK.
Fill a clean reagent reservoir with the liquid to be dispensed.
1. Dip the tip under the surface of the liquid in the reservoir and press the
trigger. This action will fi ll the tip.
2. Withdraw the tip from the liquid touching it against the edge of the reservoir
to remove excess liquid.
3. Deliver the preset volume by pressing the trigger. The display shows the
number of steps left.
4. Continue dispensing by repeating step 3. After last step a text BLOWOUT is
displayed.
5. To empty the tip completely, press trigger and hold it down.
6. Release the trigger.
If necessary, change the tip and continue pipetting.
Note
The pipette can be emptied at any time by pressing CANCEL (left
selection key)
With DILUTE function dispensing of two selected volume is possible.
Choose the DILUTE function as described above. First volume (VOL 1) is shown
on display.
Select the pipetting fi rst volume simply by pressing the scroll key up or down.
Accept the volume with OK. Next select the second volume (VOL 2) with scroll
key and accept with OK.
Optionally select the speed. Press SPEED and the speed in starts to blink.
Select the speed with scroll key and accept with OK. Select the speed out with
scroll key and accept with OK.
Stepper (multi
dispensing)
Dilute