10
Thermo Scientifi c Novus SCP/MCP User Manual
Thermo Fisher Scientifi c
Choose the PIPETTE function as described above.
Select the pipetting volume simply by pressing the scroll key up or down.
Accept the volume with OK. Optionally press SPEED and the speed in starts to
blink. Select the speed with scroll key and accept with OK. Select the speed out
with scroll key and accept with OK.
1. Dip the tip under the surface of the liquid in the reservoir and press the
trigger. The liquid is drawn in to the tip.
2. Wait until the liquid is not moving in the tip and withdraw the tip from the
liquid touching it against the edge of the reservoir to remove excess liquid.
3. To deliver the liquid press the trigger and hold it down. The blow-out is also
included to empty the tip.
4. Release the trigger to return to the ready position.
If necessary, change the tip and continue pipetting.
The Menu Map
Pipette (Forward
technique)
PIPETTE
RPIPET
STEPPER
DILUTE
PROGRAM
OPTIONS
PIPETTE
MIX+PIPETTE
COUNT
RPIPET
STEPPER
SEQ+STEPPER
DILUTE
MIX+DILUTE
MANUAL
SEQ+ASPIRATE
CALIBRATE
SERVICE
LANGUAGE
NAME
POWER
BUZZER
VERSION
BACKLIGHT
PROG 1
PROG 2
.
.
.
PROG 9
TWO POINTS
ONE POINT
COUNTER