4
Elite ORP Tester Operating Instructions
User Reset
Reset the ORP calibration to the default using the user reset function.
1. Press
to enter setup window. Scroll down by pressing
to select Reset. Press
to select reset. The
display shows User Reset and Factory Reset.
2. Press
to select User Reset. The display automatically shows No
and
Yes.
3. Scroll down by pressing to toggle between No
and
Yes
.
4. Press
to confirm either No or Yes. The display shows the user reset option with a .
Factory Reset
Reset the tester to the factory default using the factory reset function.
1. Press to enter setup window. Scroll down by pressing the
to select Reset. Press
to select Reset. The
display shows User Reset and Factory Reset.
2. Scroll down by pressing the
to toggle between the resets. Press
to select Fact Reset. The display
automatically shows No and yes.
3. Scroll down by pressing to toggle between No and Yes.
4. Press
to confirm either No or Yes. The display shows the factory reset option with a .
Manual Temperature Setting
The MTS range is 0°C to 60.0°C (32.0°F to 140.0°F). User reset will set temperature to default value 25°C or 77°F.
1. Press to select setup window. Scroll down by pressing
to select Temp Set. Press
to select Temp set.
The display shows Set °C/°F and Temp Cal.
2. Scroll down by pressing to toggle between Set °C/°F and Temp Cal. Press
to select Temp Cal.
3. The lower display shows the current measured temperature reading based on the last set offset and the upper
display shows the current measured temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the in built temperature sensor to stabilize.
5. Press
to adjust the temperature value or press the
to confrim the calibrated value as new temperature
value of the solution.
Note:
To exit without confirming the calibration, press
.
Temp Set
Settings
Key Info
Scroll
Accept
EXIT
User Reset
No
Yes
Reset
Factory Reset
Settings
Reset
Key Info
Scroll
Accept
EXIT
Temp Cal
23.4°C
23.4°C
23.4°C
Temp Set
Set °C/°F
Temp Cal
Temp Set
Settings
Key Info
Scroll
Accept
EXIT
Factory Reset
No
Yes
User Reset
Reset