Installing the Basic System
Wiring
3-8 AutoPILOT PRO Flow Computer User Guide
Thermo Fisher Scientific
Figure 3–6. Mounting the flow computer enclosure on 2” pole (metal enclosure
shown)
This section provides wiring instructions for the flow computer and the
external battery (10–30 Vdc power supply).
Warning
Ensure power is off and the area is non-hazardous before making
any connections.
Warning
Use wires suitable for at least 90°C.
Warning
Installation must be carried out in accordance with local site
requirements and regulations.
1. Remove the shipping plugs from the cable entry holes in both
enclosures. Open the doors of both enclosures.
2. Connect the ground stud on the flow computer enclosure to earth
ground before making any other connections. The ground stud is
located on the bottom of the enclosure and is marked with the ground
symbol. Use a minimum of 12 AWG stranded wire and a length as
short as possible.