17
Fault Finding
FAULT
ACTION
Foil loading tool will not pass through the rear
clamp.
Air supply not connected. Connect air to open
clamp.
Foil will not pull through.
Pulling the foil loading tool too fast or insufficient
slack fed off the roll.
Foil not transported across in clean-up cut.
Foil not trimmed square after pulling through.
Remove waste and reset under front clamp.
Failure of any air operated part to move properly
or stops during operation.
Check air supply is at the correct pressure and
that there is sufficient volume. If no improvement
call for service.
Fails to respond to an input.
Check air as above.
Doesn’t cut properly.
Tool setting problem – call for service.
Foil not picked up.
Check display for vacuum error. Check air supply.
If no improvement call for service.
Vacuum error.
Check air supply.
Doesn’t seal properly.
Time or temperature setting incorrect.
Poor seal.
Unsuitable plate. No plate support or incorrect
plate support.
Punch hits plate support with 384 PCR plate or
no plate.
Check and, if necessary, reset rear shuttle stop.
External fuse blows.
Replace and if it fails again call for service.
No power.
Check fuses. If OK, call for service.
The display shows ‘thermocouple error’
The ambient temperature is too low for the unit
to operate.