13
Communication Protocol
9.16 - To read back sealing time:
PC sends:
D<cr>
ALPS 3000 replies:
##<cr>
Where ## is a 2 digit decimal number represented in ASCII in units of 10ths of a second i.e. the string
25<cr> means the current sealing time is 2.5 seconds.
9.17 - To read back sealing temperature actual value:
PC sends:
F<cr>
ALPS 3000 replies:
###<cr>
Where ### is a 3 digit decimal number represented in ASCII, i.e. the string 170<cr> means the actual value
is 170° C.
Содержание ALPS 3000
Страница 1: ...ThermoScientificALPS3000 AutomatedLaboratory PlateSealer UserManual...