Installing the External Contact Interface Board
30
Agilent 1200 Series Getting Connected Guide for LC Devices
Thermo Scientific
To install the external contact interface board
1. For the selected module (autosampler or pump), turn off the power switch and unplug
the power supply cord (
Figure 23.
Power switches on the front of the Agilent 1200 Series modules
2. Ensure that you wear ESD protection.
Refer to the LC modules’ manuals for additional safety information and for information
about preventing ESD damage caused by an electrical discharge.
3. Using a slotted screwdriver, remove the back cover plate over the empty board slot.
4. Insert the contact interface board into the slot, and then tighten the two screws to secure
the board to the enclosure (
).
This completes the installation of the contact interface board.
CAUTION
To prevent personal injury caused by an electric shock, always turn off the
Agilent module and unplug its power supply cord from the electrical outlet before
removing the cover.
On
position
Off
position
CAUTION
To prevent damage to an instrument, always use ESD protection when
handling electronic boards and components.