2
Setting Up the LAN Interface
Setting the LAN Initialization Mode to “Using Stored”
16
Agilent 1200 Series Getting Connected Guide for LC Devices
Thermo Scientific
Setting the LAN Initialization Mode to “Using Stored”
After you configure the IP address by using the Telnet session, set the initialization mode to
use the stored IP address setting. The initialization mode is set with three of the eight
configuration switches located on the LAN board.
• On-board LAN interface—Use switches 6, 7, and 8 located on the back of the module
that has the LAN connector.
• Agilent G1369 LAN card—Use switches 4, 5, and 6 located on the LAN card.
Follow one of these procedures:
•
To set the on-board LAN interface’s initialization mode to use the stored IP address
•
To set the LAN card’s initialization mode to use the stored IP address
To set the on-board LAN interface’s initialization mode to use the stored IP address
Set the initialization switches on the back panel (
) as follows:
The On position for the toggle switch is up and the Off position is down.
To set the LAN card’s initialization mode to use the stored IP address
1. For the module that contains the LAN card, turn off the power switch and unplug its
power supply cord.
2. If an Ethernet cable connects to the Ethernet (LAN) port, disconnect the cable.
3. Ensure that you wear ESD protection.
4. Remove the LAN card from the module.
Switch #
6
7
8
Position
Off
On
Off
CAUTION
To prevent personal injury caused by an electric shock, always turn off the
module and unplug the power supply cord from the electrical outlet before removing
the cover.
CAUTION
To prevent damage to an instrument, always use ESD protection when
handling electronic boards and components.