Setting Up the LAN Interface
Setting the Stored TCP/IP Address for the LAN Interface
Thermo Scientific
Agilent 1200 Series Getting Connected Guide for LC Devices
15
9. After exiting the Telnet session, restore the TCP/IP address for the data system computer
network card (labeled Surveyor MS) to the original settings as follows (
):
a. Open the Internet Protocol Version 4 (TCP/IPv4) Properties dialog box (see
on
).
Figure 14.
Internet Protocol Version 4 (TCP/IPv4) Properties dialog box
b. In the IP Address box, type
172.16.0.101
.
c.
In the Subnet Mask box, type
255.255.0.0
.
d. Click
OK
.
e.
Exit the Control Panel.
After setting the stored IP address and subnet mask for the Agilent LAN interface, you are
ready to change the LAN interface initialization switches so that the LAN interface uses the
stored IP address. See
Setting the LAN Initialization Mode to “Using Stored”
.