12A Water Still 6-3
Thermo Scientific
Section 6
Operation
When one of the disposable demineralizer cartridges expires, the red
"CHARGE CARTRIDGE" light will come on and the water still will shut
down. Replace the cartridge as follows:
1. Turn off still and water supply.
2. Open the cartridge drain pinch clamp located behind the cartridges.
Knobs on the cartridge holders may be loosened to speed up draining.
3. Remove and discard left hand cartridge.
4. Remove right hand cartridge and install in left hand holder. Tighten
cartridge holder knob.
5. Place new cartridge upright in the right hand holder and tighten
cartridge holder knob. Close drain pinch clamps.
6. Proceed through Steps 4 - 10 in the operating instructions. Flow
meters should not need adjustment.
Your MEGA-PURE 12A Water Still is protected against overheating by
two thermal switches, located in the black "V" shaped bracket at the left
hand side of the boiler, that will shut off both the water supply and power
to the heaters when activated. Always allow the unit to cool completely (5
- 15 minutes) before resuming operation. Cool water on hot glass could
cause boiler damage. When unit has cooled, press "WATER" and
OPERATE" switches to restore normal operation. Check boiler
occasionally for proper water level.
Caution
Always allow the unit to cool completely before resuming
operation. Cool water on hot glass could cause boiler damage.
s
Demineralizer
Cartridge Changing
High Temperature
Cutoff Switch
Deionizer Cartridges
Description
Use
Part #
High Purity Cartridge
Used for high purity water
D4000377
High Capacity Cartridge
Removes ionized solids for extended capacity
D400499
Содержание 678 Series
Страница 7: ......
Страница 13: ......
Страница 22: ...12A Water Still 4 7 Thermo Scientific Section 4 Assembly Figure 4 7 Locations Figure 4 8 Installed ...
Страница 23: ......
Страница 25: ...5 2 12A Water Still Thermo Scientific Section 5 Installation Figure 5 1 Connections with Dimensions ...
Страница 29: ......
Страница 39: ......
Страница 43: ......
Страница 44: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...