15. Your MEGA-PURE 12A Water Still is now ready to be moved to its
final location. Visually check for loose connections.
16. Move the assembled water still to its final location. Refer to
Installation Instructions, and bench or wall mount the water still.
Select Method A or B to connect the water supply lines to your MEGA-
PURE 12A Water Still.
Method A
- Single supply of untreated tap water available to feed water
still - see Figure 4-6.
a. Locate "T" connector (1) and threaded connector (2) in the parts
box.
b. Connect the "T" connector (1) to the two high pressure hoses T.
These hoses connect to solenoid valves N and may be shortened to
a suitable length. Hand tighten nuts on "T" connector.
c. Thread connector (2) into your water supply and tighten. Connect
length of high pressure hose between "T" connector (1) and
threaded connector (2). Hand tighten nuts on both connectors.
d. Route 1/2" I.D. vinyl tubing to vented drain.
12A Water Still 4-5
Thermo Scientific
Section 4
Assembly
Plumbing Details
Figure 4-6.
Single Supply of Untreated Tap Water
Detail A.
Locations
Содержание 678 Series