Before troubleshooting can begin, the problem component must be
determined. That is accomplished as follows:
1. Disconnect the demineralizer and Automatic Collection System jacks
from the still control box and attempt to run the still. If still fails, see
"Still Troubleshooting." If still runs, go on to the next step.
2. Connect the demineralizer input jack back up to the still control box
and attempt to run the still. If still fails to run see "Demineralizer
Troubleshooting." If still runs, see "Automatic Collection System
Troubleshooting."
12A Water Still 8-3
Thermo Scientific
Section 8
Troubleshooting
Electrical
Still Troubleshooting
Problem
Causes
Solution
Heaters and water will not stay on.
Hi-temp cut-off switch open.
Remove control box and black "V" shaped switch bracket and
replace both switches.
Heater not working.
Heater burned out.
Replace heater (unit can be operated on less than four heaters)
Water will not come on, but light
works.
Solenoid valve not working
Check plug for connection. Replace solenoid valve.
No lights in switches.
Lights burned out.
Replace switches (switches will operate without lights)
Main breaker opens.
Heater shorted
Check heater (19 ohms for 208V or 22 ohms for 240V). Replace.
Unit off, but water still flowing.
Wrong voltage
Check inpout voltage against heater rating. Correct.
Unit off, but water still flowing.
Solenoid valve installed backwards.
Solenoid labeled "in" on input side. Check.
Содержание 678 Series
Страница 7: ......
Страница 13: ......
Страница 22: ...12A Water Still 4 7 Thermo Scientific Section 4 Assembly Figure 4 7 Locations Figure 4 8 Installed ...
Страница 23: ......
Страница 25: ...5 2 12A Water Still Thermo Scientific Section 5 Installation Figure 5 1 Connections with Dimensions ...
Страница 29: ......
Страница 39: ......
Страница 43: ......
Страница 44: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...