Operation
Settings
Thermo Scientific
43iQ Instruction Manual
3-45
●
Board Communication:
Displays OK/Fail for board communication
status.
●
Power Supply:
Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
2.5V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
●
3.3V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
●
24V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Step Board 1:
Displays current voltage readings. Alarm
limits are not changeable.
●
24V Step Board 1:
Displays current voltage readings. Alarm
limits are not changeable.
●
5.0V Step Board 2:
Displays current voltage readings. Alarm
limits are not changeable.
●
24V Step Board 2:
Displays current voltage readings. Alarm
limits are not changeable.
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......