Operation
Calibration
3-20
43iQ Instruction Manual
Thermo Scientific
The Adjust Span Coefficient screen is used to manually adjust the span
coefficient.
The following screen is shown in single range mode and dual or auto range
mode. In dual or auto range modes, “High Range” or “Low Range” is
displayed to indicate the calibration of the high or low coefficient. The
Adjust High Range Span Coefficient and Adjust Low Range Span
Coefficient screens function the same way.
Home Screen>Calibration>Advanced>Manual Calibration>Adjust
Span Coefficient (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust High Range Span Coefficient (dual or auto range
mode)
The Adjust Span Coefficient screen contains the following information:
●
Adjust Span Coefficient:
User manually adjusts span coefficient when in
single range mode.
●
Adjusted Concentration:
Read only. Shows adjusted concentration based
on adjusted span coefficient when in single range mode.
Adjust Span Coefficient
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......