Servicing
Permeation Oven Replacement
7-44
43iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the permeation oven.
Equipment required:
Phillips drive, #2
Tube release tool (optional)
CAUTION
Allow oven to cool down prior to servicing.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
2.
Unplug DMC cables as shown.
3.
Disconnect plumbing.
Figure 7–50.
Perm Oven pt 1
Permeation Oven
Replacement