Troubleshooting
Troubleshooting Guide
Thermo Scientific
43iQ Instruction Manual
6-5
Problem
Possible Cause
Action
Perform a leak test as
described on page
Reduced response or no
response to sample gas
with no alarm(s) indicated
Instrument is not properly
calibrated
Verify the SO
2
background and
SO
2
coefficient are set
appropriately in the Calibration
screen.
Instrument is not drawing in
sample as expected
Check sample flow and
pressure readings in the
Settings>Health Check>Status
and Alarms>Flow and Pressure
screen.
Use an independent flow meter
to check flows at the sample
inlet and exhaust bulkheads
(they should match).
Perform a leak test as
described on page
Check the external plumbing for
leaks or other problems.
Check all external plumbing
and the source of the sample to
verify that the SO
2
is not being
absorbed by the sampling
system. Lines carrying SO
2
must be made from clean PTFE
or stainless steel.
Detection circuit failure
Run the optical span text to
verify PMT and associated
electronics.
PMT failure
Check the PMT voltage. Adjust
if needed.
Flash lamp failure
Check the lamp voltage.
Perform a detector Autocal. If
outside of spec, replace the
detector.
Span calibration coefficient
outside acceptable limits of
0.5–2.0.
Bad span gas
Verify quality of span gas.
System leak
Perform a leak test as
described on page
Insufficient calibrator flow
Verify calibrator is providing a
flow of at least 0.8 LPM.
Excessive noise or spikes
on analog outputs
Defective or low sensitivity
Check PMT voltage and run an
optical span test. Replace PMT
with known-good unit if
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......