Optional Equipment
Connecting External Devices
Thermo Scientific
43iQ Instruction Manual
9-7
Use the following procedure to calibrate the output channel to low level.
This analog output calibration procedure reflects the zero calibration for
analog output voltage for demonstration purposes. To calibrate the 4 mA
current calibration, follow the same procedure, by selecting the 4 mA
current calibration option.
Note
This adjustment should only be performed by an instrument service
technician.
▲
1.
From the Home screen, choose
Settings>Communications>Analog
I/O>Analog Out Calibration.
2.
Depending on the output type being used, select either Analog Out
Zero Calibration (Voltage) or Analog Out 4.000 mA Calibration
(Current).
3.
A confirmation screen is presented. Select Continue to proceed with
the calibration or Return to Previous Screen.
Analog Output Zero
Calibration