Maintenance
The 915FS can perform routine maintenance. In “Maintenance”, the operator can
calibrate arc, set motor drive, calibrate motors, adjust screen, replace electrodes, and
stabilize electrodes.
Arc Calibration
Used to calibrate arc function to obtain the best splicing results when a splicing session
is first started or if the splicing environment has changed.
1. Place splicer in Splice Mode #1 (AUTO).
2. Select “Arc Calibration.”
3. Prepare and load fibers.
4. Press
8
.
5. If too low or high, load new fibers.
6. Press
8
.
7. Repeat steps 1–4 until “arc adjust OK” displays on splicer screen.