background image

CL-B101D40HZ Manual (d0098)

Page 7

Operational Display

Sub-menu 
MODE

STEP A  

Calibration

Mode

STEP B  

Calibration

Mode

 STEP E

Calibrate 

Analog 

Output 

Lo 

STEP F

Calibrate 

Analog 

Output 

Hi 

Will only appear if the analog

output option is installed

To Digital Span Selection

for Bargraph Display

See Page 5

To Two Point Digital

Calibration Mode

See Page 4

Operational Display

TWO POINT ANALOG OUTPUT

RANGE SETTING AND CALIBRATION

+

x

x

4.00

+

x

x

20.00

Determine if the Analog Output Selection Header is in the 4 to 20mA (0-20mA) posi-

tion or the 0 to 10VDC position.  If necessary, the module may have to be removed 

and the header position changed (see Component Layout below).
Note: Always disconnect power from the meter before removing the analog output mod

-

ule to adjust the mA or Volts output selection header and reinstalling it.  When power is 

reconnected, the meter’s software will automatically detect the presence or absence of 

the analog output module.

STEP A  Enter the Calibration Mode

 

1)  Press the 

P

 and 

 buttons at the same time. Display toggles between [cAL] 

and [oFF].

 

2)  Press the   or 

 button. Display changes from [oFF] to [on].

 

3)  Press the 

P

  button.  Display  toggles  between  [cAL]  and  [out]  input 

calibration.

Note: If at this point the display skips directly to toggle between Zero and the previ

-

ous Zero setting, the software is detecting that the optional analog output hardware 

is NOT installed.  

STEP B  Enter  the  Two  Point  Analog  [ouT]  Output  Range  Setting  and  Calibration 

Mode

 

1)  Press the 

P

  button.  Display  toggles  between  [cLo]  and  an  internal 

scale factor.

STEP E  Set or Calibrate [cLo] the Low Analog Value of the Analog Output Range

 

1)  Connect a multimeter to analog output pins 17 and 18 (see Rear Panel 

Pinouts on page 10). Using the   and   buttons, adjust the analog out-

put to the desired low value as measured on the multimeter. cLo may be 

adjusted to any value from –0.3 mA to 18 mA (mA output selected) or from 

–0.6 V to 8 V (volt output selected).  However, the output of cLo must 

always be less than the value selected for chi.  If a reversed analog output 

is  desired,  the  values  selected  to  establish  the  Digital  Span  can  be 

reversed (see top of page 6).  For digital readings outside the Digital Span 

selected, the analog output will not go any lower than the calibrated value 

set for cLo.  However, the analog output will linearly rise above the value 

set  for  chi,  up  to  the  the  maximum  analog  output  capability  (see  chi 

below).

 

2)  Press the 

P

  button.  Display  toggles  between  [chi]  and  an  internal 

scale factor.

STEP F  Set or Calibrate [chi] the High Analog Value of the Analog Output Range

 

1)  Using the   and   buttons, adjust the analog output to the desired 

high value as measured on the multimeter display. chi may be adjusted to 

any value from 18 mA to 24 mA (mA output) or from 8 V to 10.3 V (volt 

output). However, the value must be higher than the value selected for cLo.  

For digital readings outside the Digital Span selected, the analog output 

will linearly rise above the value set for chi, up to the maximum analog 

output capability.

 

2)  Press the 

P

 button. The meter exits the calibration mode and returns 

to the operational display.

Note: The analog output range established by the values selected for cLo and chi will 

occur, automatically scaled, between the two digital values selected for AnHi and AnLo.  

However, the analog output can linearly rise above the chi value set for digital readings 

outside the digital span selected.  See Digital Span Selection on page 6.

Two Point Analog Output Range Setting and Calibration

Two Point Analog Output Range Setting and Calibration

STEP I  Select the Range

 

1)  Using the   and   buttons, select the required range.  

    

      There are three ranges of 99.99Hz, 999.9Hz and 9999Hz

 

2)  Press the 

P

 button. The display exits the calibration  

  m o d e and returns to the operational display.

The Display/Bargraph settings are now complete. 

Two Point Analog Output Range Setting and Calibration

Range Selection Mode

Operational Display

From Bargraph Center Point

Display Mode Selection

on page 6

RANGE SELECTION

Range

[99•99]

[9999]

[999•9]

Содержание CL-B101D40HZ

Страница 1: ...Mode Selection 6 Bargraph Color Programming Mode 9 Case Dimensions 13 Component Layout 11 Connector Pinouts and Pin Descriptions 10 11 Controls and Indicators 2 Decimal Point and Brightness Selection...

Страница 2: ...irection indicated by the arrow When an alternative dotted line is shown this indicates that an alternative logic branch will be followed when a particular option is present When two buttons are shown...

Страница 3: ...See Page 5 Goes directly to Zero setting if Analog Output is not installed 4 00 20 00 Range 99 99 9999 999 9 Select one of three Ranges 99 99Hz 999 9Hz or 9999Hz PROGRAMMABLE OFFSET AND SCALE FACTOR...

Страница 4: ...6 meter may be rescaled without applying an external signal by changing the Offset and Scale factor Offset is the reading that the meter will display for a zero input The Offset may be set to any valu...

Страница 5: ...this point the display skips directly to STEP G and toggles between Cto and oFF the software is detecting that the optional analog output hardware is NOT installed Digital Span Selection For Bargraph...

Страница 6: ...h and AnLo analog low can be reversed to provide a 20 to 4mA output The digital span selected can be as small as two counts when using the analog output to function as a Control or Alarm Driver Small...

Страница 7: ...volt output selected However the output of cLo must always be less than the value selected for chi If a reversed analog output is desired the values selected to establish the Digital Span can be reve...

Страница 8: ...ress the P button Display toggles between PUM and on or oFF STEP F Select Pump PUM on or oFF 1 Using the and buttons select the Pump to be ON or OFF When PUM is selected ON and SP2 is set at a value h...

Страница 9: ...argraph Color when the bar is ABOVE SP2 Setpoint 1 Using the and buttons select the desired bargraph color grn oran or red 2 Press the P button Display toggles between CSP3 and the previous color sett...

Страница 10: ...mp Form A relays An analog output module is also shown as installed The CL B101D40HZ uses plug in type screw terminal connectors for all input and output connections The power supply connec tions pins...

Страница 11: ...A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules with 4 Isolated 5A Form A Relays Output Carrier Board Output Carrier Board Relay Module Relay Module Input Signal Conditioner Analog Output Module Main...

Страница 12: ...ditioning Modules COUPLING INPUT SIGNAL FILTER MAG NAMUR SOURCE SINK LOAD AC 20KHz 2KHz 200Hz OFF DC MAG AC LOGIC GND 24V INPUT High Accuracy Digital Measurement NPN Open Collector Proximity Switch No...

Страница 13: ...t the meter from external noise Input output leads should be routed as far away as possible from contactors control relays transformers and other noisy components Shielding cables for input output lea...

Страница 14: ...y 50 POWER SUPPLY PS1 85 265VAC 95 300VDC N C PS2 18 48VAC 10 72VDC 40 INPUT MODULES Partial List See www texmate com Unless otherwise specified Texmate will ship all modules precalibrated with factor...

Отзывы: