STEP A  Enter the Calibration Mode

 

1)  Press the 

P

 and   buttons at the same time. Display 

toggles between [CAL] and [oFF].

 2) 

Press the   or   button. Display changes from [oFF] to 

[on].

 3) 

Press the 

P

 button. Display toggles between [CAL] and  

 [out].

Note:

 If at this point, the display skips directly to STEP C and toggles 

between [ScAL] and the previous [ScAL] setting, the software 

is detecting that the optional analog output hardware is NOT 

installed.

STEP B  Select Two Point Digital Calibration of Input Signal

 

1)  Press the   or   button to select CAL [iP] for input signal 

calibration.

 

2)  Press the 

P

 button. Display toggles between [oFFS] and the  

  previous offset setting.

STEP C  Set the Offset on the Digital Display

 

1)  Using the   and   buttons, adjust the digital display to 

the desired offset. This is the reading that the meter will dis-

play for a zero input

 

2)  Press the 

P

 button. Display toggles between [ScAL] and 

the previous Scale factor.

STEP D  Set the Scale factor on the Digital Display

 1) 

Using the   and   buttons, adjust the meter display to  

the  desired Scale factor. The default value is 2000, for which a 

2V input will read 2000. If the scale factor is changed the  

display will change proportionately. Therefore if the Scale  

factor is changed to 1000 then for the same 2V input the dis 

play would read 1000.

 

2)  Press the 

P

 button. 

The Digital Rescaling is now complete

.

 

If the Digital Rescaling was successfully completed, the 

menu branches to the Digital Span Selection for Bargraph 

Display (see page 6), and the display flashes [bhi] and the 

previous setting.

The 

CL-B101D40HZ Manual (d0096)

 meter may be rescaled without applying an external signal by changing the Offset and Scale 

factor. 
Offset is the reading that the meter will display for a zero input. The Offset may be set to any value from -1999 to +9999. The 

default value of the Offset is 000 
Scale factor is the gain of the meter. The displayed reading is directly proportional to the Scale factor. The default value of the 

scale factor is 2000, but it may be set to any value between -1999 and +9999.
For an input of 2V a calibrated meter will read 2000 with the default Scale factor of 2000, 3000 with a Scale factor of 3000 and 

500 with a Scale factor of 500
If a linear scale is represented by mx + b, then the Scale Factor corresponds to the slope ‘m’ and the Offset corresponds to 

the intercept ‘b’
The internal Signal Span is limited to 3 V DC between – 1 V DC to + 2 V DC. Outputs from an Input Signal Conditioning module 

that exceed these limits will cause the meter to indicate overrange.

Note

: Most input signal conditioners have provisions for analog calibration and scaling. If the meter’s digital Scale Factor is 

set to   2000 and Offset set to 0000 then, any pre-calibrated signal condi-

tioner with an output that does not exceed – 1 V to + 2 V, will read correct-

ly in the meter without any further calibration. 

Digital Rescaling

Digital Rescaling Procedure

MAIN MENU

Operational Display

Sub-menu 
MODE

STEP A  Calibration

Mode

STEP B  Calibration

Mode

STEP C  Offset

STEP D  Scale

Factor

To Digital Span Selection for

Bargraph Display Page 6

cAL out/iP branch will only

appear if the analog output

option is installed, otherwise

cAL on goes directly to ZEro

setting

DIGITAL

RESCALING MODE

To Digital Span Selection

for Bargraph Display

See Page 7

To Two Point Analog

Output Range Setting

and Calibration

See Page 8

Содержание CL-B101D40HZ

Страница 1: ...Mode Selection 6 Bargraph Color Programming Mode 9 Case Dimensions 13 Component Layout 11 Connector Pinouts and Pin Descriptions 10 11 Controls and Indicators 2 Decimal Point and Brightness Selection...

Страница 2: ...irection indicated by the arrow When an alternative dotted line is shown this indicates that an alternative logic branch will be followed when a particular option is present When two buttons are shown...

Страница 3: ...See Page 5 Goes directly to Zero setting if Analog Output is not installed 4 00 20 00 Range 99 99 9999 999 9 Select one of three Ranges 99 99Hz 999 9Hz or 9999Hz PROGRAMMABLE OFFSET AND SCALE FACTOR...

Страница 4: ...6 meter may be rescaled without applying an external signal by changing the Offset and Scale factor Offset is the reading that the meter will display for a zero input The Offset may be set to any valu...

Страница 5: ...this point the display skips directly to STEP G and toggles between Cto and oFF the software is detecting that the optional analog output hardware is NOT installed Digital Span Selection For Bargraph...

Страница 6: ...h and AnLo analog low can be reversed to provide a 20 to 4mA output The digital span selected can be as small as two counts when using the analog output to function as a Control or Alarm Driver Small...

Страница 7: ...volt output selected However the output of cLo must always be less than the value selected for chi If a reversed analog output is desired the values selected to establish the Digital Span can be reve...

Страница 8: ...ress the P button Display toggles between PUM and on or oFF STEP F Select Pump PUM on or oFF 1 Using the and buttons select the Pump to be ON or OFF When PUM is selected ON and SP2 is set at a value h...

Страница 9: ...argraph Color when the bar is ABOVE SP2 Setpoint 1 Using the and buttons select the desired bargraph color grn oran or red 2 Press the P button Display toggles between CSP3 and the previous color sett...

Страница 10: ...mp Form A relays An analog output module is also shown as installed The CL B101D40HZ uses plug in type screw terminal connectors for all input and output connections The power supply connec tions pins...

Страница 11: ...A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules with 4 Isolated 5A Form A Relays Output Carrier Board Output Carrier Board Relay Module Relay Module Input Signal Conditioner Analog Output Module Main...

Страница 12: ...ditioning Modules COUPLING INPUT SIGNAL FILTER MAG NAMUR SOURCE SINK LOAD AC 20KHz 2KHz 200Hz OFF DC MAG AC LOGIC GND 24V INPUT High Accuracy Digital Measurement NPN Open Collector Proximity Switch No...

Страница 13: ...t the meter from external noise Input output leads should be routed as far away as possible from contactors control relays transformers and other noisy components Shielding cables for input output lea...

Страница 14: ...y 50 POWER SUPPLY PS1 85 265VAC 95 300VDC N C PS2 18 48VAC 10 72VDC 40 INPUT MODULES Partial List See www texmate com Unless otherwise specified Texmate will ship all modules precalibrated with factor...

Отзывы: