79
8-9-5. Clear and reset statuses
It is possible to cause clearing or resetting in the PSF Series power supply unit
by issuing commands or executing specific operations.
IFC (Interface clear)
Responses to the universal command IFC are as shown below:
・
The specified talker or listener status is cleared.
・
The GP-IB buffer, output queue and standing-by command remain
unchanged.
・
The SRQ remains unchanged.
・
The remote status and LLO setting remain unchanged.
・
Panel setting remains unchanged.
DCL/SDC
(
Device clear
)
Responses to the universal command DCL and address command SDC are
as shown below:
・
The GP-IB buffer, output queue and standing-by command are cleared.
・
The interface status (specified talker or listener status) is cleared.
・
The SRQ and status byte are cleared.
・
The remote status and LLO setting are cleared and the power supply unit
enters the local state.
・
Panel setting remains unchanged.
*
RST
(
Reset command
)
Responses to the IEEE488.2 command *RST are as shown below:
・
The specified talker or listener status is cleared.
・
The GP-IB buffer, output queue and standing-by command are cleared.
・
The SRQ, status byte and mask setting are cleared.
・
The remote status and LLO setting remain unchanged.
・
Panel setting remains unchanged.
Turning on power again
When power is turned on again, the statuses are as shown below:
・
The specified talker or listener status is cleared.
・
The GP-IB buffer, output queue and standing-by command are cleared.
・
The SRQ, status byte and mask setting are cleared.
・
The unit enters the local status.
LLO set before turning off power is cleared.
・