5-5-2. OVP/OCP function (02)
●
OVP: (Over Voltage Protection)
The OVP function turns off output when the output voltage of the PSF Series
exceeds the preset OVP value.
The setting range is from 10.0V to 840.0V. The resolution is 1V.
●
OCP: (Over Current Protection)
The OCP function turns off output when the output current of the PSF Series
exceeds the preset OCP value. The setting range is from 0.10A to
3.15A(400H)
、
6.30A(800H). The resolution is 0.01A. The resolution is
different in one-control parallel operation.
◆
See Section
“2. SPECIFICATIONS”
.
Operation procedure
1. Press the MENU key until menu
number “02” is displayed.
2. Every press of the
ENTER/CHECK key alternated
OVP and OCP.
3.
When an intended function is
displayed, rotate the encoder
to set an intended value
4. Press the ENTER/CHECK key
to validate it.
V
W
A
Hi-
Ω
W
OFF TIMER
OUTPUT
*
ON/
・
OFF
CV
CC
CP
V
A
W
MENU ENTER
ESC
CHECK
DISP
5. press the ESC/DISP key to
return to the nomal state.