14. Output setting of reference B
Command: SETB X1
Function: Set the output of the reference B. This command is
available only when the maximum value of the
reference B is set with the MODEB command. In
other condition, this command causes a GP-IB run
error. The reference output is set to zero when the
GP-600B is turned on.
Parameter: X1: Output value
(0 to maximum set value of the reference B)
The format of the numeral is XXXX.XX.
15. Output setting of reference C
Command: SETC X1
Function: Set the output of the reference C. This command is
available only when the maximum value of the
reference C is set with the MODEC command. In
other condition, this command causes a GP-IB run
error. The reference output is set to zero when the
GP-600B is turned on.
Parameter: X1: Output value
(0 to maximum set value of the reference C)
The format of the numeral is XXXX.XX.
16. Output setting of reference D
Command: SETD X1
Function: Set the output of the reference D. This command is
available only when the maximum value of the
reference D is set with the MODED command. In
other condition, this command causes a GP-IB run
error. The reference output is set to zero when the
GP-600B is turned on.
Parameter: X1: Output value
(0 to maximum set value of the reference D)
The format of the numeral is XXXX.XX.