30
8. Output Off channel setting
Command: OFFCH X1
Function:
Specify channel(s) to which the GP-600B turns off
output (i.e., it resets the output voltage and current to
zero and turns off output) in the case where an OVP,
alarm, or power switch Off condition occurs in the
power supply units. It is set to 1 when the GP-600B is
turned on.
Parameter: X1: 0- Do not turn off output.
1- Turn off output of only the channel where an OVP,
alarm, or power switch Off condition occurs.
2- Turn off outputs of both channels 1 and 2.
9.
Maximum value setting of reference A
Command: MODEA X1
Function:
If the maximum value of the reference A is set with this
command, the reference output may be only changed
with the SETA command. The reference output is
reset to zero if this command is sent while the
reference is being output.
Parameter: X1: Maximum output (0.01 to 9999.99)
The format of the numeral is XXXX.XX.
10. Maximum value setting of reference B
Command: MODEB X1
Function: If the maximum value of the reference B is set with this
command, the reference output may be only changed
with the SETB command. The reference output is
reset to zero if this command is sent while the
reference is being output.
Parameter: X1: Maximum output (0.01 to 9999.99)
The format of the numeral is XXXX.XX.