33
If the parameter of the AMP, VOLT, OUT, SETA, SETB, SETC or
SETD command is omitted, the parameter is automatically set to the
value used previously. This function allows the previous setting to
be restored even when the OUTPUT OFF key is pressed and each
output is turned off.
Example: When the rated values (references A and B) of channel 1
are set with the MODE command and the rated values
(maximum values) of the references C and D are set with
the MODEC and MODED commands
SELECT 1 :VOLT :AMP :OUT
SETC :SETD :SELECT 2 :OUT
17.
SRQ masking time setting
Command: MTIME X1
Function: Specify the time of masking service request output in
CV or CC operation when the output is turned on (with
the OUT 1 command).
The time is set in 10 ms multiplied by the parameter
value. (10 ms x parameter value)
This command only masks the service requests caused
in CV and CC operations of a channel whose output is
turned on. Service requests due to other causes are
issued. The parameter of this command is always set
to 10 (i.e., 100 ms) when the GP-600B is turned on.
Parameter: X1: 2 to 100
The format of the numeral is XXXX.XX.
NOTE: The GP-600B checks the status of the power supply units at
the 100ms interval. Thus, it may issue a service request in
CV or CC operation after turning on output a maximum of 100
ms later than the preset masking time.