Control()/CONTROL_STATUS(): 0x00 and 0x01
57
SLUUBE8 – September 2018
Copyright © 2018, Texas Instruments Incorporated
Data Commands
4.1.27 ALERT_SET: 0x0068
This command forces the ALERT pin to be set, no matter the conditions. The
PIN_CONTROL_EN
command must be run to enable this feature.
4.1.28 ALERT_RESET: 0x0069
This command forces a reset on the ALERT pin. The
PIN_CONTROL_EN
command must be run to
enable this feature.
4.1.29 EDV_Threshold: 0x0073
This MAC subcommand returns the values of the EDV Threshold.
4.1.30 EXIT_ CAL: 0x0080
This instructs the fuel gauge to exit CALIBRATION mode.
4.1.31 ENTER_CAL: 0x0081
This instructs the fuel gauge to enter CALIBRATION mode and reset
AnalogCount()
to 0 if
OperationStatus()[CALMD]
is set.
[CALMD]
is controlled by the
CAL_MODE()
command. Voltage must be
above 2.9 V for accurate calibration.
4.1.32 ENTER_CFG_UPDATE: 0x0090
Instructs the fuel gauge to set the
OperationStatus()[DFGUPDATE]
bit to 1 and enter CONFIG UPDATE
mode. This command is only available when the fuel gauge is UNSEALED.
NOTE:
To read the flag, the host must wait at least 2 seconds.
4.1.33 EXIT_CFG_UPDATE_REINIT: 0x0091
This command instructs the fuel gauge to exit CONFIG UPDATE mode and the gauge is reinitialized.
4.1.34 EXIT_CFG_UPDATE: 0x0092
This command instructs the fuel gauge to exit CONFIG UPDATE mode and the gauge is not reinitialized.
4.2
Temperature(): 0x06 and 0x07
This read- and write-word function returns an unsigned integer value of the temperature in units of 0.1°K
measured by the fuel gauge. See
Temperature Measurement Options,
and
Table 4-6. Temperature Measurement Options
Operation
Config A
[TEMPS]
Temperature() Read Command
Temperature() Write Command
0
Returns internal temperature as read from an internal sensor. This data
is also available using the
InternalTemperature()
function.
The data is ignored.
1
Returns external temperature read from an external thermistor.
X
Returns the
Temperature()
value previously written.
Sets the
Temperature()
to be used for gauging
calculations.