5 Parameterizing, adjusting and analyzing software (P2A software)
70
Field/button
Explanation
Default value
Freely definable output value for the
respective type of analog output (V or
mA), 1 decimal place.
[Activate]
The entered default value is forwarded
to the corresponding analog output
and to the test contacts by clicking.
A warning informs that the value is
being transmitted to the connected
instrument in the event of existing
cabling.
Now check the analog output using a
precise multimeter.
[Deactivate]
Finish entering the electrical variables
at the analog output.
The analog output returns to the
current reading again.
3.
Click on
[OK]
or
[Cancel]
to close the dialogue.
- The analog output and the relay return to Measuring Mode
again.