4 Transmitter
50
Shown on the
display
5
Can be used for
collective alarm
6
Additional
message end
7
Analog adjustment
x
T ambient high
x
x
T ambient low
x
x
Supply voltage low
x
x
Watchdog error
x
Perform the
Confirm message
function (acknowledgement of the
alarm via the control keys on the transmitter):
• The message/alarm is no longer shown on the display and the
optical alarm goes out, where applicable. If multiple
messages/alarms are active at the same time, all are reset
simultaneously.
• If at least one message is assigned to the collective alarm, the
collective alarm is reset. If the collective alarm is set on a relay,
the relay is also reset, meaning switched to its neutral position.
4.5.5.
Namur fault conditions
If the faults named in the following table occur, the analog outputs
output special values that enable a general fault warning in the
higher-level control system. The values correspond to the "Namur"
industry standard.
Status
message in
the display
Class
Display
value in the
display
Analog output
0 to
20 mA
4 to
20 mA
1 V
5 V 10 V
Watchdog
error
Error
Previous
value stops
21 mA
3.8 mA 1.1 V 5.5 V 11 V
Value below
min. scale
Underrange Reading
0 mA
3.8 mA 0 V
0 V 0 V
Value above
max. scale
Overrange Reading
20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
Pressure too
high
Overrange ooooo ·
20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
Содержание 6351