5 Parameterizing, adjusting and analyzing software (P2A software)
68
2.
Select the command
Delete
in the context menu.
- The instrument or parameter file is deleted from the list.
5.3.2.6.
Creating a new instrument file
It is possible to create an instrument file without restarting the P2A
software.
✓
Transmitter must be connected.
1.
Click on
File > New connection
in the menu bar.
-
Connection to the transmitter is established.
5.3.3.
Analyzing/testing the transmitter
In this section, you can test the outputs of the connected
instrument, read off the limit values and reset the parameters to the
factory settings.
The function is only available for instrument files.
5.3.3.1.
Analyzing/testing the instrument
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Perform action:
Action
Explanation
Carrying out factory
reset
Reset the unit, limit value and hysteresis
parameters to factory settings (see below).
Testing analog
output
Test channel 1 (see 4.4.6.4. Testing
functionality of analog outputs).
Test switch outputs Manually switch relays 1 to 4 to test for
proper function (see 4.4.6.4. Testing
functionality of relay outputs).
Display min./max.
values
Overview of the minimum and maximum
values measured since the last reset of the
transmitter (see 5.3.3.5. Displaying min./max.
values).
3.
Click on
[OK]
or
[Cancel]
to close the dialogue.
5.3.3.2.
Carrying out factory reset
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.