Terumo Azur CX 35 Скачать руководство пользователя страница 9

- 4 -

Figure 4 

– Position of Marker Bands for Detachment 

To minimize the potential risk of aneurysm or vessel perforation

DO NOT

 advance the distal end of the delivery system past the

distal tip of the microcatheter.

DETACHMENT OF THE COIL

 

32. The Azur Detachment Controller is pre-loaded with battery power

and will activate when a delivery pusher is properly connected.  It
is in a “power off” mode when no delivery pusher is attached.  It is
not necessary to push the button on the side of the Azur
Detachment Controller to activate it.

33. Verify that the RHV is firmly locked around the delivery pusher

before attaching the Azur Detachment Controller to ensure that
the coil does not move during the connection process.

34. Although the delivery pusher’s gold connectors are designed to

be compatible with blood and contrast, every effort should be
made to keep the connectors free of these items.  If there appears
to be blood or contrast on the connectors, wipe the connectors
with sterile water or saline solution before connecting to the Azur
Detachment Controller.

35. Connect the proximal end of the delivery pusher to the Azur

Detachment Controller by firmly inserting the proximal end of the
delivery pusher into the funnel section of the Azur Detachment
Controller.  See Figure 2.

36. When the Azur Detachment Controller is properly connected to

the delivery pusher, a single audible tone will sound and the light
will turn green to signal that it is ready to detach the coil.  If the
detachment button is not pushed within 30 seconds, the solid
green light will slowly flash green.  Both flashing green and solid
green lights indicate that the device is ready to detach.  If the
green light does not appear, check to ensure that the connection
has been made.  If the connection is correct and no green light
appears, replace the Azur Detachment Controller.

37. Verify the coil position before pushing the detachment button.
38. Push the detachment button.  When the button is pushed, an

audible tone will sound and the light will flash green.

39. At the end of the detachment cycle, three audible tones will sound

and the light will flash yellow three times.  This indicates that the
detachment cycle is complete.  If the coil does not detach during
the detachment cycle, leave the Azur Detachment Controller
attached to the delivery pusher and attempt another detachment
cycle when the light turns green.

40. The light will turn red after the number of detachment cycles

specified on the Azur Detachment Controller labeling.  DO NOT
use the Azur Detachment Controller if the light is red.  Discard the
Azur Detachment Controller and replace it with a new one when
the light is red.

41. Verify detachment of the coil by first loosening the RHV valve,

then pulling back slowly on the delivery system and verifying that
there is no coil movement.  If the implant did not detach, do not
attempt to detach it more than two additional times.  If it does not
detach after the third attempt, remove the delivery system.

42. After detachment has been confirmed, slowly retract and remove

the delivery pusher. 

Advancing the delivery pusher once the

coil has been detached involves the risk of aneurysm or 
vessel rupture. Do NOT advance the delivery pusher once the 
coil has been detached.

 

43. Verify the position of the coil angiographically through the guide

catheter.

44. Additional coils may be deployed into the lesion as described

above.  Prior to removing the microcatheter from the treatment
site, place an appropriately sized guidewire completely through

the microcatheter lumen to ensure that no part of the last coil
remains within the microcatheter.

The physician has the discretion to modify the coil deployment
technique to accommodate the complexity and variation in
embolization procedures.  Any technique modifications must be
consistent with the previously described procedures, warnings,
precautions and patient safety information.

SPECIFICATIONS FOR AZUR DETACHMENT CONTROLLER 

Output voltage: 8 ± 1 VDC

Cleaning, preventative inspection, and maintenance: The
Azur Detachment Controller is a single use device,
preloaded with battery power, and packaged sterile.  No
cleaning, inspection, or maintenance is required.  If the
device does not perform as described in the Detachment
section of these Instructions, discard the Azur Detachment
Controller and replace it with a new unit.

The Azur Detachment Controller is a single use device.  Do
not reuse, reprocess or resterilize. Reuse, reprocessing or
resterilization may compromise the structural integrity of the
device and/or lead to device failure which, in turn, may result
in patient injury, illness, or death. Reuse, reprocessing, or
resterilization may also create a risk of contamination of the
device and/or cause patient infection or cross-infection,
including, but not limited to, the transmission of infectious
disease(s) from one patient to another. Contamination of the
device may lead to injury, illness or death of the patient.

Batteries are pre-loaded into the Azur Detachment
Controllers.  Do not attempt to remove or replace the
batteries prior to use.

After use, dispose of the Azur Detachment Controller in a
manner consistent with local regulations.

PACKAGING AND STORAGE 

The Azur system is placed inside a protective, plastic dispenser
hoop and packaged in a pouch and unit carton.  The Azur system
and dispenser hoop will remain sterile unless the package is
opened, damaged, or the expiration date has passed.  Store at a
controlled room temperature in a dry place.

The Azur Detachment Controller is packaged separately in a
protective pouch and carton.  The Azur Detachment Controller
has been sterilized; it will remain sterile unless the pouch is
opened, damaged, or the expiration date has passed. Store at a
controlled room temperature in a dry place.

SHELF LIFE 

See the product label for the device shelf life.  Do not use the
device beyond the labeled shelf life.

MR INFORMATION

 

The coil implant has been determined to be 

MR conditional

according to the terminology specified in the American Society for
Testing and Materials (ASTM) International, Designation: F2503-
08.
Non-clinical testing demonstrated that the coil implant is

 MR

conditional

. A patient can be scanned safely, immediately after

placement under the following conditions:

 

Static magnetic field of 3 Tesla or less

 

Maximum spatial gradient field of 720 Gauss/cm or less

MRI-Related Heating
In non-clinical testing, the coil implant produced a maximum
temperature rise of 1.7°C during MRI performed for 15 minutes of

Содержание Azur CX 35

Страница 1: ...gle use only Do not reuse reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in...

Страница 2: ...timate the size of the lesion to be treated 9 For aneurysm occlusion the diameter of the first and second coils should never be less than the width of the aneurysm neck or the propensity for the coils...

Страница 3: ...otating the delivery pusher may result in a stretched coil or premature detachment of the coil from the delivery pusher which could result in coil migration Angiographic assessment should also be perf...

Страница 4: ...ction including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may lead to injury illness or death of the patient Batteries are pre...

Страница 5: ...age or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability or responsibility in...

Страница 6: ...use reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in patient injury illnes...

Страница 7: ...will allow for fluoroscopic road mapping during the procedure 6 Select a microcatheter with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guid...

Страница 8: ...the introducer sheath Excessive tightening could damage the device 23 Push the coil into the lumen of the microcatheter Use caution to avoid catching the coil on the junction between the introducer s...

Страница 9: ...e risk of aneurysm or vessel rupture Do NOT advance the delivery pusher once the coil has been detached 43 Verify the position of the coil angiographically through the guide catheter 44 Additional coi...

Страница 10: ...not limited to any implied warranties of merchantability or fitness for particular purpose Handling storage cleaning and sterilization of the device as well as factors relating to the patient diagnos...

Страница 11: ...ation of the device and or cause patient infection or cross infection including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may...

Страница 12: ...e may be increased For vessel occlusion it is suggested that the diameter of the initial coil placed be slightly larger than the actual vessel diameter to prevent displacement or migration For aneurys...

Страница 13: ...catheter tip from moving away from the intended delivery location 20 Once the coil has been deployed and prior to removing the delivery catheter advance the guidewire through the catheter tip to verif...

Страница 14: ...sequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability o...

Страница 15: ...device and or lead to device failure which in turn may result in patient injury illness or death Reuse reprocessing or resterilization may also create a risk of contamination of the device and or caus...

Страница 16: ...er with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guidewire COIL SIZE SELECTION 7 Perform fluoroscopic road mapping 8 Measure and estimate...

Страница 17: ...ppropriately sized coil Movement of the coil may indicate that the coil could migrate once it is detached DO NOT rotate the delivery pusher during or after delivery of the coil into the vasculature Ro...

Страница 18: ...of the patient Batteries are pre loaded into the Azur Detachment Controllers Do not attempt to remove or replace the batteries prior to use After use dispose of the Azur Detachment Controller in a man...

Страница 19: ...not be liable for any incidental or consequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for...

Страница 20: ...PD110323 Rev A 1 MicroVention Inc Instructions for Use Detachable...

Страница 21: ...sitioned in the vessel or aneurysm within the specified reposition time from the time the device is first introduced into the microcatheter If the coil cannot be positioned and detached within this ti...

Страница 22: ...ation replace the device If the green light remains solid green for the entire three second observation continue using the device 15 Hold the device just distal to the shrink lock and pull the shrink...

Страница 23: ...The coil implant has been determined to be MR conditional according to the terminology specified in the American Society for Testing and Materials ASTM International Designation F2503 08 Non clinical...

Отзывы: