Terumo Azur CX 35 Скачать руководство пользователя страница 8

- 3 -

Figure 2 

– Azur Detachment Controller 

13. Remove the Azur Detachment Controller from its protective

packaging. Pull the white pull-tab from the side of the detachment
controller. Discard the pull-tab and place the detachment
controller in the sterile field.  The Azur Detachment Controller is
packaged separately as a sterile device.  

Do not use any power

source other than the Azur Detachment Controller to detach
the coil.  The Azur Detachment Controller is intended to be
used on one patient.  Do not attempt to re-sterilize or
otherwise re-use the Azur Detachment Controller.

14. Prior to using the device, remove the proximal end of the delivery

pusher from the packaging hoop.  Use care to avoid
contaminating this end of the delivery pusher with foreign
substances such as blood or contrast.  Firmly insert the proximal
end of the delivery pusher into the funnel section of the Azur
Detachment Controller.  See Figure 2.  

Do not push the

detachment button at this time.

15. Wait three seconds and observe the indicator light on the

detachment controller.

If the green light does not appear or if a red light appears,
replace the device.

If the light turns green, then turns off at any time during the
three-second observation, replace the device.

If the green light remains solid green for the entire three-
second observation, continue using the device.

16. Hold the device just distal to the shrink-lock and pull the shrink-

lock proximally to expose the tab on introducer sheath.  See
Figure 3.

Figure 3 - Pull Shrink Lock Proximally 

17. Slowly advance the coil out of the introducer sheath and inspect

the coil for any irregularities or damage.  

If any damage to the

coil or delivery pusher is observed, DO NOT use the device.

18. If pre–softening of the Azur implant is desired, advance it out of

the distal end of the introducer sheath and immerse it in warm
sterile saline or warm lactated Ringer’s injection.  Alternatively,
hold it in a flow of steam until it curls, usually about five to ten
seconds.  When using steam, appropriate sterile technique should
be used.  In addition, the Azur system may be used without pre-
softening.

19. With the distal end of the introducer sheath pointed downward

and the implant still in the warm saline, warm lactated Ringer’s

injection or flow of steam, gently retract the implant back
completely into the introducer sheath about 1 to 2 cm.

INTRODUCTION AND DEPLOYMENT OF THE AZUR SYSTEM 

20. Open the RHV on the microcatheter just enough to accept the

introducer sheath of the Azur system.

21. Insert the introducer sheath of the Azur system through the RHV.

Flush the introducer until it is completely purged of air and saline
flush exits the proximal end.

22. Seat the distal tip of the introducer sheath at the distal end of the

microcatheter hub and close the RHV 

lightly

 around the

introducer sheath to secure the RHV to the introducer.

Do not over-tighten the RHV around the introducer sheath.
Excessive tightening could damage the device.

23. Push the coil into the lumen of the microcatheter.  Use caution to

avoid catching the coil on the junction between the introducer
sheath and the hub of the microcatheter.  

Initiate timing using a

stopwatch or timer at the moment the device enters the
microcatheter.  Detachment must occur within the specified
reposition time.

24. Push the Azur system through the microcatheter until the proximal

end of the delivery pusher meets the proximal end of the
introducer sheath.  Loosen the RHV.  Retract the introducer
sheath just out of the RHV.  Close the RHV around the delivery
pusher.  Slide the introducer sheath completely off of the delivery
pusher.  Use care not to kink the delivery system.  To prevent
premature hydration of the Azur system, ensure that there is flow
from the saline flush.

25. Discard the introducer sheath.  The Azur system cannot be re-

sheathed after introduction into the microcatheter.

26. At this time, fluoroscopic guidance should be initiated.  Depending

on length of microcatheter used, fluoroscopy initiation may be
delayed to minimize exposure.

27. Under fluoroscopic guidance, slowly advance the coil out the tip of

the microcatheter.  Continue to advance the coil into the lesion
until optimal deployment is achieved.  Reposition if necessary.  If
the coil size is not suitable, remove and replace with another
device.  If undesirable movement of the coil is observed under
fluoroscopy following placement and prior to detachment, remove
the coil and replace with another more appropriately sized coil.
Movement of the coil may indicate that the coil could migrate once
it is detached.  

DO NOT

 rotate the delivery pusher during or after

delivery of the coil into the vasculature.  Rotating the delivery
pusher may result in a stretched coil or premature detachment of
the coil from the delivery pusher, which could result in coil
migration.  Angiographic assessment should also be performed
prior to detachment to ensure that the coil mass is not protruding
into undesired vasculature.

28. Complete the deployment and any repositioning so that the coil

will be detached within the reposition time specified in Table 1.
After the specified time, the swelling of the hydrophilic polymer
may prevent passage through the microcatheter and damage the
coil.  

If the coil cannot be properly positioned and detached

within the specified time, simultaneously remove the device
and the microcatheter

.

29. Advance the coil into the desired site until the radiopaque marker

on the delivery pusher is aligned or slightly distal of microcatheter
distal tip RO marker, positioning the detachment zone just outside
the microcatheter tip.  See Figure 4.

30. Tighten the RHV to prevent movement of the coil.
31. Verify repeatedly that the distal shaft of the delivery pusher is not

under stress before coil detachment.  Axial compression or
tension could cause the tip of the microcatheter to move during
coil delivery.  Catheter tip movement could cause the aneurysm
or vessel to perforate.

Pull shrink-lock 
proximally 

Tab on introducer
sheath 

Содержание Azur CX 35

Страница 1: ...gle use only Do not reuse reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in...

Страница 2: ...timate the size of the lesion to be treated 9 For aneurysm occlusion the diameter of the first and second coils should never be less than the width of the aneurysm neck or the propensity for the coils...

Страница 3: ...otating the delivery pusher may result in a stretched coil or premature detachment of the coil from the delivery pusher which could result in coil migration Angiographic assessment should also be perf...

Страница 4: ...ction including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may lead to injury illness or death of the patient Batteries are pre...

Страница 5: ...age or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability or responsibility in...

Страница 6: ...use reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in patient injury illnes...

Страница 7: ...will allow for fluoroscopic road mapping during the procedure 6 Select a microcatheter with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guid...

Страница 8: ...the introducer sheath Excessive tightening could damage the device 23 Push the coil into the lumen of the microcatheter Use caution to avoid catching the coil on the junction between the introducer s...

Страница 9: ...e risk of aneurysm or vessel rupture Do NOT advance the delivery pusher once the coil has been detached 43 Verify the position of the coil angiographically through the guide catheter 44 Additional coi...

Страница 10: ...not limited to any implied warranties of merchantability or fitness for particular purpose Handling storage cleaning and sterilization of the device as well as factors relating to the patient diagnos...

Страница 11: ...ation of the device and or cause patient infection or cross infection including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may...

Страница 12: ...e may be increased For vessel occlusion it is suggested that the diameter of the initial coil placed be slightly larger than the actual vessel diameter to prevent displacement or migration For aneurys...

Страница 13: ...catheter tip from moving away from the intended delivery location 20 Once the coil has been deployed and prior to removing the delivery catheter advance the guidewire through the catheter tip to verif...

Страница 14: ...sequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability o...

Страница 15: ...device and or lead to device failure which in turn may result in patient injury illness or death Reuse reprocessing or resterilization may also create a risk of contamination of the device and or caus...

Страница 16: ...er with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guidewire COIL SIZE SELECTION 7 Perform fluoroscopic road mapping 8 Measure and estimate...

Страница 17: ...ppropriately sized coil Movement of the coil may indicate that the coil could migrate once it is detached DO NOT rotate the delivery pusher during or after delivery of the coil into the vasculature Ro...

Страница 18: ...of the patient Batteries are pre loaded into the Azur Detachment Controllers Do not attempt to remove or replace the batteries prior to use After use dispose of the Azur Detachment Controller in a man...

Страница 19: ...not be liable for any incidental or consequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for...

Страница 20: ...PD110323 Rev A 1 MicroVention Inc Instructions for Use Detachable...

Страница 21: ...sitioned in the vessel or aneurysm within the specified reposition time from the time the device is first introduced into the microcatheter If the coil cannot be positioned and detached within this ti...

Страница 22: ...ation replace the device If the green light remains solid green for the entire three second observation continue using the device 15 Hold the device just distal to the shrink lock and pull the shrink...

Страница 23: ...The coil implant has been determined to be MR conditional according to the terminology specified in the American Society for Testing and Materials ASTM International Designation F2503 08 Non clinical...

Отзывы: