background image

- 3 -

Detachment Controller.  See Figure 2.  

Do not push the 

detachment button at this time.

 

14. Wait three seconds and observe the indicator light on the

detachment controller.

If the green light does not appear or if a red light appears,
replace the device.

If the light turns green, then turns off at any time during the
three-second observation, replace the device.

If the green light remains solid green for the entire three-
second observation, continue using the device.

15. Hold the device just distal to the shrink-lock and pull the shrink-

lock proximally to expose the tab on introducer sheath.  See
Figure 3.

Figure 3 - Pull Shrink Lock Proximally 

16. Slowly advance the coil out of the introducer sheath and inspect

the coil for any irregularities or damage.  

If any damage to the

coil or delivery pusher is observed, DO NOT use the device.

17. With the distal end of the introducer sheath pointed downward,

gently retract the implant back completely into the introducer
sheath about 1 to 2 cm.

INTRODUCTION AND DEPLOYMENT OF THE AZUR SYSTEM 

18. Open the RHV on the catheter just enough to accept the

introducer sheath of the Azur system.

19. Insert the introducer sheath of the Azur system through the RHV.

Flush the introducer until it is completely purged of air and saline
flush exits the proximal end.

20. Seat the distal tip of the introducer sheath at the distal end of the

catheter hub and close the RHV 

lightly

 around the introducer

sheath to secure the RHV to the introducer.

Do not over-tighten the RHV around the introducer sheath.
Excessive tightening could damage the device.

21. Push the coil into the lumen of the catheter.  Use caution to avoid

catching the coil on the junction between the introducer sheath
and the hub of the catheter.  

Initiate timing using a stopwatch

or timer at the moment the device enters the catheter.
Detachment must occur within the specified reposition time.

22. Push the Azur system through the catheter until the proximal end

of the delivery pusher meets the proximal end of the introducer
sheath.  Loosen the RHV.  Retract the introducer sheath just out
of the RHV.  Close the RHV around the delivery pusher.  Slide the
introducer sheath completely off of the delivery pusher.  Use care
not to kink the delivery system.  To prevent premature hydration
of the Azur system, ensure that there is flow from the saline flush.

23. Discard the introducer sheath.  The Azur system cannot be re-

sheathed after introduction into the microcatheter.

24. At this time, fluoroscopic guidance should be initiated.  Depending

on length of catheter used, fluoroscopy initiation may be delayed
to minimize exposure.

25. Under fluoroscopic guidance, slowly advance the coil out the tip of

the catheter.  Continue to advance the coil into the lesion until
optimal deployment is achieved.  Reposition if necessary.  If the
coil size is not suitable, remove and replace with another device.
If undesirable movement of the coil is observed under fluoroscopy
following placement and prior to detachment, remove the coil and
replace with another more appropriately sized coil.  Movement of

the coil may indicate that the coil could migrate once it is 
detached.  

DO NOT

 rotate the delivery pusher during or after 

delivery of the coil into the vasculature.  Rotating the delivery 
pusher may result in a stretched coil or premature detachment of 
the coil from the delivery pusher, which could result in coil 
migration.  Angiographic assessment should also be performed 
prior to detachment to ensure that the coil mass is not protruding 
into undesired vasculature. 

26. Complete the deployment and any repositioning so that the coil

will be detached within the reposition time specified in Table 1.
After the specified time, the swelling of the hydrophilic polymer
may prevent passage through the catheter and damage the coil.

If the coil cannot be properly positioned and detached within
the specified time, simultaneously remove the device and the
catheter

.

27. Advance the coil into the desired site until the radiopaque marker

on the delivery pusher is aligned or slightly distal of catheter distal
tip RO marker, positioning the detachment zone just outside the
catheter tip.  See Figure 4.

28. Tighten the RHV to prevent movement of the coil.
29. Verify repeatedly that the distal shaft of the delivery pusher is not

under stress before coil detachment.  Axial compression or
tension could cause the tip of the catheter to move during coil
delivery.  Catheter tip movement could cause the aneurysm or
vessel to perforate.

Figure 4 

– Position of Marker Bands for Detachment 

DETACHMENT OF THE COIL

 

30. The Azur Detachment Controller is pre-loaded with battery power

and will activate when a delivery pusher is properly connected.  It
is in a “power off” mode when no delivery pusher is attached.  It is
not necessary to push the button on the side of the Azur
Detachment Controller to activate it.

31. Verify that the RHV is firmly locked around the delivery pusher

before attaching the Azur Detachment Controller to ensure that
the coil does not move during the connection process.

32. Although the delivery pusher’s gold connectors are designed to

be compatible with blood and contrast, every effort should be
made to keep the connectors free of these items.  If there appears
to be blood or contrast on the connectors, wipe the connectors
with sterile water or saline solution before connecting to the Azur
Detachment Controller.

33. Connect the proximal end of the delivery pusher to the Azur

Detachment Controller by firmly inserting the proximal end of the
delivery pusher into the funnel section of the Azur Detachment
Controller.  See Figure 2.

34. When the Azur Detachment Controller is properly connected to

the delivery pusher, a single audible tone will sound and the light
will turn green to signal that it is ready to detach the coil.  If the
detachment button is not pushed within 30 seconds, the solid
green light will slowly flash green.  Both flashing green and solid
green lights indicate that the device is ready to detach.  If the
green light does not appear, check to ensure that the connection
has been made.  If the connection is correct and no green light
appears, replace the Azur Detachment Controller.

35. Verify the coil position before pushing the detachment button.
36. Push the detachment button.  When the button is pushed, an

audible tone will sound and the light will flash green.

37. At the end of the detachment cycle, three audible tones will sound

and the light will flash yellow three times.  This indicates that the
detachment cycle is complete.  If the coil does not detach during
the detachment cycle, leave the Azur Detachment Controller

Pull shrink-lock 
proximally 

Tab on introducer 
sheath 

Содержание Azur CX 35

Страница 1: ...gle use only Do not reuse reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in...

Страница 2: ...timate the size of the lesion to be treated 9 For aneurysm occlusion the diameter of the first and second coils should never be less than the width of the aneurysm neck or the propensity for the coils...

Страница 3: ...otating the delivery pusher may result in a stretched coil or premature detachment of the coil from the delivery pusher which could result in coil migration Angiographic assessment should also be perf...

Страница 4: ...ction including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may lead to injury illness or death of the patient Batteries are pre...

Страница 5: ...age or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability or responsibility in...

Страница 6: ...use reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in patient injury illnes...

Страница 7: ...will allow for fluoroscopic road mapping during the procedure 6 Select a microcatheter with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guid...

Страница 8: ...the introducer sheath Excessive tightening could damage the device 23 Push the coil into the lumen of the microcatheter Use caution to avoid catching the coil on the junction between the introducer s...

Страница 9: ...e risk of aneurysm or vessel rupture Do NOT advance the delivery pusher once the coil has been detached 43 Verify the position of the coil angiographically through the guide catheter 44 Additional coi...

Страница 10: ...not limited to any implied warranties of merchantability or fitness for particular purpose Handling storage cleaning and sterilization of the device as well as factors relating to the patient diagnos...

Страница 11: ...ation of the device and or cause patient infection or cross infection including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may...

Страница 12: ...e may be increased For vessel occlusion it is suggested that the diameter of the initial coil placed be slightly larger than the actual vessel diameter to prevent displacement or migration For aneurys...

Страница 13: ...catheter tip from moving away from the intended delivery location 20 Once the coil has been deployed and prior to removing the delivery catheter advance the guidewire through the catheter tip to verif...

Страница 14: ...sequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability o...

Страница 15: ...device and or lead to device failure which in turn may result in patient injury illness or death Reuse reprocessing or resterilization may also create a risk of contamination of the device and or caus...

Страница 16: ...er with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guidewire COIL SIZE SELECTION 7 Perform fluoroscopic road mapping 8 Measure and estimate...

Страница 17: ...ppropriately sized coil Movement of the coil may indicate that the coil could migrate once it is detached DO NOT rotate the delivery pusher during or after delivery of the coil into the vasculature Ro...

Страница 18: ...of the patient Batteries are pre loaded into the Azur Detachment Controllers Do not attempt to remove or replace the batteries prior to use After use dispose of the Azur Detachment Controller in a man...

Страница 19: ...not be liable for any incidental or consequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for...

Страница 20: ...PD110323 Rev A 1 MicroVention Inc Instructions for Use Detachable...

Страница 21: ...sitioned in the vessel or aneurysm within the specified reposition time from the time the device is first introduced into the microcatheter If the coil cannot be positioned and detached within this ti...

Страница 22: ...ation replace the device If the green light remains solid green for the entire three second observation continue using the device 15 Hold the device just distal to the shrink lock and pull the shrink...

Страница 23: ...The coil implant has been determined to be MR conditional according to the terminology specified in the American Society for Testing and Materials ASTM International Designation F2503 08 Non clinical...

Отзывы: