Terumo Azur CX 35 Скачать руководство пользователя страница 11

- 1 -

Azur

 

Peripheral

 

Coil System 

Helical HydroCoil

®

 Embolization System 

(Pushable) 

Instructions for Use 

DEVICE DESCRIPTION 

The Pushable Azur Peripheral Coil System (Azur system) consists of a
coil implant packaged in a coil introducer along with an introducer stylet.
The coil is platinum-based with an outer layer of hydrogel polymer.
The Azur system is available in helical 18-system and helical 35-system
configurations.  Both configurations are available in a broad range of
secondary diameters and lengths to meet the needs of the physician.
Each system must be delivered through a microcatheter or catheter
within the specified ID range, using a specified guidewire size.

System 

Catheter/Microcatheter ID 

inches 

mm 

18-system

0.021 – 0.022

0.53 – 0.56

35-system

0.041 – 0.047

1.04 – 1.19

System 

Guidewire OD 

inches 

mm 

18-system

0.018

0.46

35-system

0.035

0.89

INDICATIONS FOR USE 

The Azur system is intended to reduce or block the rate of blood flow in
vessels of the peripheral vasculature.  It is intended for use in the
interventional radiologic management of arteriovenous malformations,
arteriovenous fistulae, aneurysms, and other lesions of the peripheral
vasculature.
This device should only be used by physicians who have undergone
training in the use of the Azur system for embolization procedures as
prescribed by a representative from Terumo

 

or a Terumo-authorized

distributor.

CONTRAINDICATIONS 

Use of the Azur system is contraindicated in any of the following
circumstances:

When superselective coil placement is not possible.

When end arteries lead directly to nerves.

When arteries supplying the lesion to be treated are not large
enough to accept emboli.

When the A-V shunt is larger than the coil.

In the presence of severe atheromatous disease.

In the presence of vasospasm (or likely onset of vasospasm).

POTENTIAL COMPLICATIONS 

Potential complications include, but are not limited to: hematoma at the
site of entry, vessel/aneurysm perforation, unintentional occlusion of the
parent artery, incomplete filling, emboli, hemorrhage, ischemia,
vasospasm, edema, coil migration or misplacement, clot formation,
revascularization, post-embolization syndrome, and neurological deficits
including stroke and possibly death.

REQUIRED ADDITIONAL ITEMS 

Appropriately-sized catheter/microcatheter (with wire reinforcement
required for delivery through tortuous vasculature)

Steerable guidewire compatible with coil system and delivery
catheter/microcatheter

Rotating hemostatic Y valve (RHV)

Sterile saline

Pressurized sterile saline drip

One-way stopcock

1cc syringe

WARNINGS AND PRECAUTIONS 

Federal law (USA) restricts this device to sale by or on the 
order of a physician. 

The Azur system is sterile and non-pyrogenic unless the unit
package is opened or damaged.

The Azur system is intended for single use only.  Do not reuse,
reprocess or resterilize. Reuse, reprocessing or resterilization may
compromise the structural integrity of the device and/or lead to
device failure which, in turn, may result in patient injury, illness, or
death. Reuse, reprocessing, or resterilization may also create a
risk of contamination of the device and/or cause patient infection or
cross-infection, including, but not limited to, the transmission of
infectious disease(s) from one patient to another. Contamination of
the device may lead to injury, illness or death of the patient.

Angiography is required for pre-embolization evaluation, operative
control, and post-embolization follow up.  Fluoroscopic
roadmapping is recommended to achieve optimal device
placement.

Always inspect the Azur system prior to both preparation and
insertion to ensure that the coil has not shifted within the introducer
or migrated into the introducer caps.  If the coil is not secure within
the introducer prior to both the preparation and introduction
processes, damage may result.

Hydration of the Azur

 

system prior to use is 

mandatory

.

  

A 3-

minute hydration period is required to soften the coil.  Failure to
hydrate may result in the coil not taking its secondary shape, which
can result in deployment away from the intended location,
migration, or protrusion outside the delivery location.

The coil must be delivered through a compatibly-sized catheter or
microcatheter with a PTFE inner surface coating using a
compatibly-sized guidewire.  Failure to correctly size the delivery
system may result in damage to the device and necessitate
removal of both the device and delivery catheter from the patient.

Always select a wire-reinforced delivery catheter/microcatheter
when delivering the coil through highly tortuous vasculature.  Non-
reinforced catheters may ovalize under such circumstances,
potentially resulting in coil damage and necessitating removal of
both the device and delivery catheter from the patient.

Do not use a syringe to deliver the coil.  The coil is intended to be
delivered using a compatible guidewire only.  Delivery via syringe
injection may result in the coil not taking its secondary shape,
which can result in deployment away from the intended location,
migration, or protrusion outside the delivery location.

Do not advance the coil with excessive force.  If unusual resistance
is noted during advancement, determine its cause before
proceeding by verifying the appropriate delivery catheter and
guidewire are being used, and that both are free from damage and
kinking.  If necessary, replace the delivery catheter, coil, and/or
guidewire before proceeding.

The coil is not retractable or repositionable.  If a coil must be
retrieved from the vasculature after deployment, do not attempt to
withdraw the coil with a retrieval device, such as a snare, into the
delivery catheter.  This could damage the coil and result in device
separation.  Remove the coil, microcatheter, and any retrieval
device from the vasculature simultaneously.

If the coil and/or pushing guidewire get stuck within the delivery
catheter lumen, do not continue advancing.  Remove the catheter,
and replace the catheter, coil, and/or guidewire when necessary.

Delivery of multiple coils is generally required to achieve the
desired occlusion of some vessels, aneurysms, and vascular
lesions.  The desired procedural endpoint is angiographic
occlusion.  The filling properties of the coil facilitate angiographic
occlusion and reduce the need to tightly pack.  Multiple
embolization procedures may be required to achieve the desired
occlusion of some vessels/vascular lesions.

Содержание Azur CX 35

Страница 1: ...gle use only Do not reuse reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in...

Страница 2: ...timate the size of the lesion to be treated 9 For aneurysm occlusion the diameter of the first and second coils should never be less than the width of the aneurysm neck or the propensity for the coils...

Страница 3: ...otating the delivery pusher may result in a stretched coil or premature detachment of the coil from the delivery pusher which could result in coil migration Angiographic assessment should also be perf...

Страница 4: ...ction including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may lead to injury illness or death of the patient Batteries are pre...

Страница 5: ...age or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability or responsibility in...

Страница 6: ...use reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in patient injury illnes...

Страница 7: ...will allow for fluoroscopic road mapping during the procedure 6 Select a microcatheter with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guid...

Страница 8: ...the introducer sheath Excessive tightening could damage the device 23 Push the coil into the lumen of the microcatheter Use caution to avoid catching the coil on the junction between the introducer s...

Страница 9: ...e risk of aneurysm or vessel rupture Do NOT advance the delivery pusher once the coil has been detached 43 Verify the position of the coil angiographically through the guide catheter 44 Additional coi...

Страница 10: ...not limited to any implied warranties of merchantability or fitness for particular purpose Handling storage cleaning and sterilization of the device as well as factors relating to the patient diagnos...

Страница 11: ...ation of the device and or cause patient infection or cross infection including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may...

Страница 12: ...e may be increased For vessel occlusion it is suggested that the diameter of the initial coil placed be slightly larger than the actual vessel diameter to prevent displacement or migration For aneurys...

Страница 13: ...catheter tip from moving away from the intended delivery location 20 Once the coil has been deployed and prior to removing the delivery catheter advance the guidewire through the catheter tip to verif...

Страница 14: ...sequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability o...

Страница 15: ...device and or lead to device failure which in turn may result in patient injury illness or death Reuse reprocessing or resterilization may also create a risk of contamination of the device and or caus...

Страница 16: ...er with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guidewire COIL SIZE SELECTION 7 Perform fluoroscopic road mapping 8 Measure and estimate...

Страница 17: ...ppropriately sized coil Movement of the coil may indicate that the coil could migrate once it is detached DO NOT rotate the delivery pusher during or after delivery of the coil into the vasculature Ro...

Страница 18: ...of the patient Batteries are pre loaded into the Azur Detachment Controllers Do not attempt to remove or replace the batteries prior to use After use dispose of the Azur Detachment Controller in a man...

Страница 19: ...not be liable for any incidental or consequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for...

Страница 20: ...PD110323 Rev A 1 MicroVention Inc Instructions for Use Detachable...

Страница 21: ...sitioned in the vessel or aneurysm within the specified reposition time from the time the device is first introduced into the microcatheter If the coil cannot be positioned and detached within this ti...

Страница 22: ...ation replace the device If the green light remains solid green for the entire three second observation continue using the device 15 Hold the device just distal to the shrink lock and pull the shrink...

Страница 23: ...The coil implant has been determined to be MR conditional according to the terminology specified in the American Society for Testing and Materials ASTM International Designation F2503 08 Non clinical...

Отзывы: