background image

PD110323 Rev. A 

MicroVention, Inc.

34.  When the Azur Detachment Controller is properly connected to the delivery 

pusher, a single audible tone will sound and the light will turn green to signal 
that it is ready to detach the coil. If the detachment button is not pushed within 
30 seconds, the solid green light will slowly flash green. Both flashing green and 
solid green lights indicate that the device is ready to detach. If the green light 
does not appear, check to ensure that the connection has been made. If the 
connection is correct and no green light appears, replace the Azur Detachment 
Controller.

35.  Verify the coil position before pushing the detachment button.
36.  Push the detachment button. When the button is pushed, an audible tone will

sound and the light will flash green.

37.

At the end of the detachment cycle, three audible tones will sound and the 
light will flash yellow three times. This indicates that the detachment cycle is 
complete. If the coil does not detach during the detachment cycle, leave the 
Azur Detachment Controller attached to the delivery pusher and attempt another 
detachment cycle when the light turns green. 

38.

The light will turn red after the number of detachment cycles specified on 
the Azur Detachment Controller labeling. DO NOT use the Azur detachment 
controller if the light is red. Discard the Azur Detachment Controller and replace 
it with a new one when the light is red.

39.

Verify detachment of the coil by first loosening the RHV valve, then pulling back 
slowly on the delivery system and verifying that there is no coil movement. If the 
implant did not detach, do not attempt to detach it more than two additional 
times. If it does not detach after the third attempt, remove the delivery system.

40.

After detachment has been confirmed, slowly retract and remove the delivery 
pusher. 

Advancing the delivery pusher once the coil has been detached 

involves the risk of aneurysm or vessel rupture. Do NOT advance the 

delivery pusher once the coil has been detached

.

41.

Verify the position of the coil angiographically through the guide catheter.

42.

Additional coils may be deployed into the lesion as described above. Prior to 
removing the microcatheter from the treatment site, place an appropriately sized 
guidewire completely through the microcatheter lumen to ensure that no part of 
the last coil remains within the microcatheter.

The physician has the discretion to modify the coil deployment technique to 
accommodate the complexity and variation in embolization procedures. Any technique 
modifications must be consistent with the previously described procedures, warnings, 
precautions and patient safety information.

SPECIFICATIONS FOR AZUR DETACHMENT CONTROLLER

•  Output voltage: 8 ± 1 VDC
•  Cleaning, preventative inspection, and maintenance: The Azur Detachment 

Controller is a single use device, preloaded with battery power, and packaged 
sterile. No cleaning, inspection, or maintenance is required. If the device does not 
perform as described in the Detachment section of these Instructions, discard the 
Azur Detachment Controller and replace it with a new unit.

•  The Azur Detachment Controller is a single use device. Do not reuse, reprocess or

resterilize. Reuse, reprocessing or resterilization may compromise the structural 
integrity of the device and/or lead to device failure which, in turn, may result 
in patient injury, illness, or death. Reuse, reprocessing, or resterilization may 
also create a risk of contamination of the device and/or cause patient infection 
or cross-infection, including, but not limited to, the transmission of infectious 
disease(s) from one patient to another. Contamination of the device may lead to 
injury, illness or death of the patient.

•  Batteries are pre-loaded into the Azur Detachment Controllers. Do not attempt to 

remove or replace the batteries prior to use.

•  After use, dispose of the Azur Detachment Controller in a manner consistent with

local regulations.

PACKAGING AND STORAGE

The Azur system is placed inside a protective, plastic dispenser hoop and packaged 
in a pouch and unit carton. The Azur system and dispenser hoop will remain sterile 
unless the package is opened, damaged, or the expiration date has passed. Store at a 
controlled room temperature in a dry place.
The Azur Detachment Controller is packaged separately in a protective pouch and 
carton. The Azur Detachment Controller has been sterilized; it will remain sterile 
unless the pouch is opened, damaged, or the expiration date has passed. Store at a 
controlled room temperature in a dry place.

SHELF LIFE

See the product label for the device shelf life. Do not use the device beyond the 
labeled shelf life.

MR INFORMATION  

 

The coil implant has been determined to be 

MR conditional

 according to the 

terminology specified in the American Society for Testing and Materials (ASTM) 
International, Designation: F2503-08. 
Non-clinical testing demonstrated that the coil implant is 

MR conditional

. A patient 

can be scanned safely, immediately after placement under the following conditions:

•  Static magnetic field of 3 Tesla or less
•  Maximum spatial gradient field of 720 Gauss/cm or less

21.

Push the coil into the lumen of the microcatheter. Use caution to avoid catching 
the coil on the junction between the introducer sheath and the hub of the 
microcatheter. 

Initiate timing using a stopwatch or timer at the moment 

the device enters the microcatheter. Detachment must occur within the 

specified reposition time.

22.  Push the Azur system through the microcatheter until the proximal end of the 

delivery pusher meets the proximal end of the introducer sheath. Loosen the 
RHV. Retract the introducer sheath just out of the RHV. Close the RHV around 
the delivery pusher. Slide the introducer sheath completely off of the delivery 
pusher. Use care not to kink the delivery system. To prevent premature hydration 
of the Azur system, ensure that there is flow from the saline flush.

23.

Discard the introducer sheath. The Azur system cannot be re-sheathed after 
introduction into the microcatheter.

24.

At this time, fluoroscopic guidance should be initiated. Depending on length of 
microcatheter used, fluoroscopy initiation may be delayed to minimize exposure.

25.

Under fluoroscopic guidance, slowly advance the coil out the tip of the 
microcatheter. Continue to advance the coil into the lesion until optimal 
deployment is achieved. Reposition if necessary. If the coil size is not suitable, 
remove and replace with another device. If undesirable movement of the coil 
is observed under fluoroscopy following placement and prior to detachment, 
remove the coil and replace with another more appropriately sized coil. 
Movement of the coil may indicate that the coil could migrate once it is 
detached. 

DO NOT

 rotate the delivery pusher during or after delivery of the 

coil into the vasculature. Rotating the delivery pusher may result in a stretched 
coil or premature detachment of the coil from the delivery pusher, which could 
result in coil migration. Angiographic assessment should also be performed 
prior to detachment to ensure that the coil mass is not protruding into undesired 
vasculature.

26.

Complete the deployment and any repositioning so that the coil will be 
detached within the reposition time specified in Table 1. After the specified 
time, the swelling of the hydrophilic polymer may prevent passage through the 
microcatheter and damage the coil. 

If the coil cannot be properly positioned 

and detached within the specified time, simultaneously remove the device 

and the microcatheter. 

27.

Advance the coil into the desired site until the radiopaque marker on the 
delivery pusher is aligned or slightly distal of microcatheter distal tip RO marker, 
positioning the detachment zone just outside the microcatheter tip. See Figure 4.

28.

Tighten the RHV to prevent movement of the coil. 

29.

Verify repeatedly that the distal shaft of the delivery pusher is not under stress 
before coil detachment. Axial compression or tension could cause the tip of the 
microcatheter to move during coil delivery. Catheter tip movement could cause 
the aneurysm or vessel to perforate.

Figure 4 – Position of Marker Bands for Detachment

Coil Implant

Proximal end of coil is aligned with 
microcatheter RO tip

Microcatheter Distal Tip RO Marker

Distal end of delivery 
pusher

Microcatheter

 To minimize the potential risk of aneurysm or vessel perforation 

DO NOT

 advance 

the distal end of the delivery system past the distal tip of the microcatheter.

DETACHMENT OF THE COIL

30.

The Azur Detachment Controller is pre-loaded with battery power and will 
activate when a delivery pusher is properly connected. It is in a “power off” 
mode when no delivery pusher is attached. It is not necessary to push the 
button on the side of the Azur Detachment Controller to activate it.

31.

Verify that the RHV is firmly locked around the delivery pusher before attaching 
the Azur Detachment Controller to ensure that the coil does not move during the 
connection process.

32.

Although the delivery pusher’s gold connectors are designed to be compatible 
with blood and contrast, every effort should be made to keep the connectors 
free of these items. If there appears to be blood or contrast on the connectors, 
wipe the connectors with sterile water or saline solution before connecting to the 
Azur Detachment Controller.

33.  Connect the proximal end of the delivery pusher to the Azur Detachment 

Controller by firmly inserting the proximal end of the delivery pusher into the 
funnel section of the Azur Detachment Controller. See Figure 2.

Содержание Azur CX 35

Страница 1: ...gle use only Do not reuse reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in...

Страница 2: ...timate the size of the lesion to be treated 9 For aneurysm occlusion the diameter of the first and second coils should never be less than the width of the aneurysm neck or the propensity for the coils...

Страница 3: ...otating the delivery pusher may result in a stretched coil or premature detachment of the coil from the delivery pusher which could result in coil migration Angiographic assessment should also be perf...

Страница 4: ...ction including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may lead to injury illness or death of the patient Batteries are pre...

Страница 5: ...age or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability or responsibility in...

Страница 6: ...use reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in patient injury illnes...

Страница 7: ...will allow for fluoroscopic road mapping during the procedure 6 Select a microcatheter with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guid...

Страница 8: ...the introducer sheath Excessive tightening could damage the device 23 Push the coil into the lumen of the microcatheter Use caution to avoid catching the coil on the junction between the introducer s...

Страница 9: ...e risk of aneurysm or vessel rupture Do NOT advance the delivery pusher once the coil has been detached 43 Verify the position of the coil angiographically through the guide catheter 44 Additional coi...

Страница 10: ...not limited to any implied warranties of merchantability or fitness for particular purpose Handling storage cleaning and sterilization of the device as well as factors relating to the patient diagnos...

Страница 11: ...ation of the device and or cause patient infection or cross infection including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may...

Страница 12: ...e may be increased For vessel occlusion it is suggested that the diameter of the initial coil placed be slightly larger than the actual vessel diameter to prevent displacement or migration For aneurys...

Страница 13: ...catheter tip from moving away from the intended delivery location 20 Once the coil has been deployed and prior to removing the delivery catheter advance the guidewire through the catheter tip to verif...

Страница 14: ...sequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability o...

Страница 15: ...device and or lead to device failure which in turn may result in patient injury illness or death Reuse reprocessing or resterilization may also create a risk of contamination of the device and or caus...

Страница 16: ...er with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guidewire COIL SIZE SELECTION 7 Perform fluoroscopic road mapping 8 Measure and estimate...

Страница 17: ...ppropriately sized coil Movement of the coil may indicate that the coil could migrate once it is detached DO NOT rotate the delivery pusher during or after delivery of the coil into the vasculature Ro...

Страница 18: ...of the patient Batteries are pre loaded into the Azur Detachment Controllers Do not attempt to remove or replace the batteries prior to use After use dispose of the Azur Detachment Controller in a man...

Страница 19: ...not be liable for any incidental or consequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for...

Страница 20: ...PD110323 Rev A 1 MicroVention Inc Instructions for Use Detachable...

Страница 21: ...sitioned in the vessel or aneurysm within the specified reposition time from the time the device is first introduced into the microcatheter If the coil cannot be positioned and detached within this ti...

Страница 22: ...ation replace the device If the green light remains solid green for the entire three second observation continue using the device 15 Hold the device just distal to the shrink lock and pull the shrink...

Страница 23: ...The coil implant has been determined to be MR conditional according to the terminology specified in the American Society for Testing and Materials ASTM International Designation F2503 08 Non clinical...

Отзывы: