Appendix A: KT-10 v2
Plus
Feature
This document is in addition to the KT-10 v2 manual. It is intended to provide
information specific to the operation of the
Plus
feature for the KT-10 v2. The procedures
outlined in this section are also applicable for conductivity measurements and calibration.
Please consult the KT-10 v2’s operations manual for general operating instructions.
The KT-10 v2
Plus
has an increased range of up to
10 SI
units for magnetic susceptibility
measurements. Direct iron ore concentration estimates can be obtained from the meter’s
display with the use of a magnetite calibration curve already loaded into the unit.
The KT-10 v2
Plus
meter has been developed to allow users to calibrate the meter to
specific ore types for quick recognition in the field. It can be used for sample selection,
core analysis and grade control. With the ability to use three different calibration curves,
one of which is the pre-installed magnetite curve, users have a flexible, user friendly
instrument suitable for several different applications. If the samples or cores have a
different composition or structure than those used to create the pre-installed calibration
curve included in the KT-10 v2
Plus
, the user can program up to two additional
calibration curves that are specific to the samples and cores they are measuring.
At the end of this guide you will find recommendations for preparing samples for
site specific calibration procedures.
The KT-10 v2
Plus
has an increased range of operation (10 SI). Users can select
different units (SI units or %) to receive measurement results in; this option is located in
the Setup menu. From the main menu select the Setup option and the following menu
will be presented.
Figure 129: Setup menu